Abstract

Atrial leadless pacemakers (ALPMs) offer a minimally invasive solution for patients requiring atrial pacing, but current designs face significant challenges related to fixation stability, perforation risk, and retrievability. This study presents a novel self-expanding nitinol fixation system designed for deployment within the left atrial appendage (LAA), incorporating a flexible adapter for secure pacemaker engagement and retrieval. Finite-element simulations were conducted to assess gravitational displacement across different anatomical orientations, and fixture-expansion behavior was analyzed under various mesh configurations. The pacemaker drop analysis results demonstrated minimal displacement in neutral and upward-tilted LAA models, with increased instability observed in downward-tilted orientations. The fixture-expansion study showed that the 0.2 mm mesh design provided adequate mechanical strength and strain tolerance while maintaining a compact profile. This novel fixation system improves current ALPM limitations by providing stable, retrievable anchoring and favorable biomechanical performance. It may also serve as a dual-function platform for atrial pacing and stroke prevention when integrated with a left atrial appendage (LAA) occluder. These findings support further preclinical validation for clinical translation.

Original languageEnglish
Article number512
JournalBioengineering
Volume12
Issue number5
DOIs
StatePublished - 12 05 2025

Bibliographical note

Publisher Copyright:
© 2025 by the authors.

UN SDGs

This output contributes to the following UN Sustainable Development Goals (SDGs)

  1. SDG 3 - Good Health and Well-being
    SDG 3 Good Health and Well-being

Keywords

  • finite-element analysis
  • fixture
  • leadless pacemaker
  • left atrial appendage (LAA)
  • nitinol alloy