Analyses and applications of single scan dose profiles in computed tomography

Hui-Yu Tsai, Chuan-Jong Tung*, M. H. Huang, Y. L. Wan

*Corresponding author for this work

Research output: Contribution to journalJournal Article peer-review

25 Scopus citations

Abstract

Comprehensive analyses and measurements of computed tomography (CT) single-scan dose profiles were performed for several scanners and operating conditions. Measurements were made using two types of thermoluminescent dosimeters, LiF:Mg,Cu,P and CaSO4:Dy, and two CT dosimetry phantoms, head and body. Analyses of CT single-scan dose profiles were made in terms of a Gaussian function for primary radiation and a Lorentzian function for scattered radiation. This function was used to investigate several common descriptions of the CT dose, including the computed tomography dose index (CTDI) and the multiple scan average dose. The relative percentage of scatter versus primary radiation to the contribution of CTDI at the central and peripheral locations was determined and analyzed. The correlation between CTDI of thermoluminescent dosimeter measurements and pencil-shaped ionization chamber measurements was determined. A method for estimating organ dose from CT was developed and compared to organ-dose estimates from Monte Carlo simulations.

Original languageEnglish
Pages (from-to)1982-1989
Number of pages8
JournalMedical Physics
Volume30
Issue number8
DOIs
StatePublished - 01 08 2003

Keywords

  • CT dose
  • CTDI
  • Computed tomography
  • MSAD