Biomechanical Comparison between Isobar and Dynamic-Transitional Optima (DTO) Hybrid Lumbar Fixators: A Lumbosacral Finite Element and Intersegmental Motion Analysis

  • Shih Hao Chen
  • , Chih Kun Hsiao
  • , Chih Wei Wang
  • , Hsiang Ho Chen
  • , Zheng Cheng Zhong*
  • *Corresponding author for this work
  • Tzu Chi University
  • E-Da Hospital
  • National Yang Ming Chiao Tung University

Research output: Contribution to journal › Journal Article › peer-review

8 Scopus citations

Abstract

Biomechanical performance of longitudinal component in dynamic hybrid devices was evaluated to display the load-transfer effects of Dynesys cord spacer or Isobar damper-joint dynamic stabilizer on junctional problem based on various disc degenerations. The dynamic component was adapted at the mildly degenerative L3-L4 segment, and the static component was fixed at the moderately degenerative L4-L5 segment under a displacement-controlled mode for the finite element study. Furthermore, an intersegmental motion behavior was analyzed experimentally on the synthetic model under a load-controlled mode. Isobar or DTO hybrid fixator could reduce stress/motion at transition segment, but compensation was affected at the cephalic adjacent segment more than the caudal one. Within the trade-off region (as a motion-preserving balance between the transition and adjacent segments), the stiffness-related problem was reduced mostly in flexion by a flexible Dynesys cord. In contrast, Isobar damper afforded the effect of maximal allowable displacement (more than peak axial stiffness) to reduce stress within the pedicle and at facet joint. Pedicle-screw travel at transition level was related to the extent of disc degeneration in Isobar damper-joint (more than Dynesys cord spacer) attributing to the design effect of axial displacement and angular rotation under motion. In biomechanical characteristics relevant to clinical use, longitudinal cord/damper of dynamic hybrid lumbar fixators should be designed with less interface stress occurring at the screw-vertebral junction and facet joint to decrease pedicle screw loosening/breakage under various disc degenerations.

Original languageEnglish
Article number8273853
JournalBioMed Research International
Volume2022
DOIs
StatePublished - 2022

Bibliographical note

Publisher Copyright:
© 2022 Shih-Hao Chen et al.

Fingerprint

Dive into the research topics of 'Biomechanical Comparison between Isobar and Dynamic-Transitional Optima (DTO) Hybrid Lumbar Fixators: A Lumbosacral Finite Element and Intersegmental Motion Analysis'. Together they form a unique fingerprint.

Cite this