Biomechanical evaluation of the effects of implant neckwall thickness and abutment screw size: A 3D nonlinear finite element analysis

Ming Dih Jeng, Yang Sung Lin, Chun Li Lin*

*Corresponding author for this work

Research output: Contribution to journalJournal Article peer-review

18 Scopus citations

Abstract

In this study, we evaluate the influence of implant neck wall thickness and abutment screw size on alveolar bone and implant component biomechanical responses using nonlinear finite element (FE) analysis. Twelve internal hexagon Morse taper implant-abutment connection FE models with three different implant sizes (diameters 4, 5, and 6 mm), secured with 1.4, 1.6, and 1.8 mm abutment screws to fit with three unilateral implant neck wall thicknesses of 0.45, 0.50, and 1.00 mm, were constructed to perform simulations. Nonlinear contact elements were used to simulate realistic interface fixation within the implant system. A 200 N concentrated force was applied toward the center of a hemispherical load cap and inclined 30ffi relative to the implant axis as the loading condition. The simulation results indicated that increasing the unilateral implant neck wall thickness from 0.45 to 1.00 mm can significantly decrease implant, abutment, and abutment screw stresses and bone strain, decreased to 58%, 48%, 54%, and 70%, respectively. Variations in abutment screw size only significantly influenced abutment screw stress, and the maximum stress dissipation rates were 10% and 29% when the diameter was increased from 1.4 to 1.6 and 1.8 mm, respectively. We conclude that the unilateral implant neck wall thickness is the major design factor for the implant system and implant neck wall thickness in effectively decreasing implant, abutment, and abutment screw stresses and bone strain.

Original languageEnglish
Article number3471
JournalApplied Sciences (Switzerland)
Volume10
Issue number10
DOIs
StatePublished - 01 05 2020
Externally publishedYes

Bibliographical note

Publisher Copyright:
© 2020 by the authors.

Keywords

  • Abutment screw
  • Biomechanics
  • Dental implant
  • Finite element analysis
  • Implant neck wall

Fingerprint

Dive into the research topics of 'Biomechanical evaluation of the effects of implant neckwall thickness and abutment screw size: A 3D nonlinear finite element analysis'. Together they form a unique fingerprint.

Cite this