Skip to main navigation Skip to search Skip to main content

Biomechanical Impact of Footprint Mismatch and Implant Positioning in Cervical Disk Replacement: A Finite Element Analysis

  • Ming Kai Hsieh
  • , Fu Cheng Kao
  • , Yu Cheng Wang
  • , Ping Yeh Chiu
  • , Yun Da Li
  • , Tsung Ting Tsai
  • , Po Liang Lai
  • , Weng Pin Chen*
  • *Corresponding author for this work
  • Chang Gung University
  • National Taipei University of Technology
  • New Taipei Municipal Tucheng Hospital (Built and Operated by Chang Gung Medical Foundation)

Research output: Contribution to journal › Journal Article › peer-review

Abstract

BACKGROUND AND OBJECTIVES: – A footprint mismatch between the cervical disk replacement implant and vertebral endplates is common; however, debate remains regarding the optimal implant position when a footprint-undersized device is unavoidable. Our purpose is to evaluate the biomechanical impact of a footprint-undersized ball-and-socket cervical disk replacement on spinal kinematics, facet joint stress, endplate contact stress, and adjacent disk stress in different sagittal position.METHODS: – A validated C4-C7 finite element model was developed from computed tomography data. A Prodisc® C Vivo prosthesis was implanted at C5-C6 in 2 configurations: (1) footprint-matched and (2) mismatched using undersized implants positioned anteriorly, centrally, and posteriorly. Physiological follower loads (98 N) and pure moments (2 Nm) were applied to simulate flexion, extension, lateral bending, and axial rotation. Ranges of motion, implant-endplate contact stress, core stress, facet joint stress, and adjacent disk stress were compared.RESULTS: – The mismatched-anterior model reduced C5-C6 ranges of motion by 22% in flexion but increased facet joint stress by 166% during extension and adjacent disk stress by up to 148% in flexion. Implant-endplate contact stress increased by up to 277% in the mismatched-anterior model, suggesting a higher risk of subsidence. The mismatched-central model exhibited lower increases in core stress, while the mismatched-posterior model preserved comparable flexion and extension motions, indicating that both models demonstrated relatively more favorable biomechanical performance.CONCLUSION: – Undersized cervical disk implants significantly alter segmental biomechanics, especially when placed anteriorly, increasing the risk of subsidence, facet overload, and adjacent segment degeneration.

Original languageEnglish
JournalOperative neurosurgery (Hagerstown, Md.)
Early online date25 03 2026
DOIs
StateE-pub ahead of print - 25 03 2026

Bibliographical note

Publisher Copyright:
© Congress of Neurological Surgeons 2026. All rights reserved.

Keywords

  • Adjacent segment degeneration
  • Biomechanics
  • Cervical disk arthroplasty
  • Finite element analysis
  • Footprint
  • Mismatch

Fingerprint

Dive into the research topics of 'Biomechanical Impact of Footprint Mismatch and Implant Positioning in Cervical Disk Replacement: A Finite Element Analysis'. Together they form a unique fingerprint.

Cite this