Abstract
This study investigates the biomechanical interaction of different mini-plate fixation types (shapes/sizes and patterns) with segmental advancement levels on the Le Fort I osteotomy using the non-linear finite element (FE) approach. Nine models were generated under a standard 1-piece LeFort I osteotomy for advancement with 3, 6 and 9 mm distances and four mini-plates with three fixation patterns including LL, LI, and II patterns placed on the maxillae models by integrating computed tomography images and computer-aided design system. The axial and oblique occlusal forces were 250 N applied to each premolar/molar and 125 N applied at 30° inclination to the tooth long axis and from palatal to buccal, respectively. The relative micro-movement values between the two maxillary bone segments and maximum mini-plate stress increased obviously with maxilla advancement increment and the increasing trend can be fitted by exponential curve. The corresponding values in II mini-plate fixation presented apparently high values in all simulated cases. The mini-plate stress concentration locations were found at the bending regions to increase high fracture risk. The mini-plate yield strength can be mapped to a critical (limited) advancement for three types of fixations for safe consideration. This study concluded that L-shaped mini-plates with lateral fixation are recommended to provide better stability. The risk for mini-plate fracture and bone relapse increases when maxillary advancement is larger than a critical value of 5 mm in the Le Fort I osteotomy.
| Original language | English |
|---|---|
| Pages (from-to) | 1704-1713 |
| Number of pages | 10 |
| Journal | Computer Methods in Biomechanics and Biomedical Engineering |
| Volume | 19 |
| Issue number | 16 |
| DOIs | |
| State | Published - 09 12 2016 |
Bibliographical note
Publisher Copyright:© 2016 Informa UK Limited, trading as Taylor & Francis Group.
Keywords
- Finite element
- Le Fort I osteotomy
- advancement
- fixation
- mini-plate