Abstract
We evaluated 887 hips in 672 patients with uncemented MicroStructured Omnifit acetabular components for liner locking complications. We found 2 types of radiographic signs of liner locking system complications in 7 hips, developing between 2 and 4 years postoperatively. The incidence of liner locking system complications was 0.8% using this modular acetabular component. We recommend that a patient who has received a total hip arthroplasty including a MicroStructured Omnifit acetabular component should be monitored frequently for radiographic signs of liner locking system complications, especially with a polyethylene thickness of less than 8 mm.
Original language | English |
---|---|
Pages (from-to) | 31-33 |
Number of pages | 3 |
Journal | Acta Orthopaedica Scandinavica |
Volume | 71 |
Issue number | 1 |
DOIs | |
State | Published - 02 2000 |