Skip to main navigation Skip to search Skip to main content

Concurrent photoacoustic-ultrasound imaging using single-laser pulses

  • Shi Yao Hung
  • , Wen Shao Wu
  • , Bao Yu Hsieh
  • , Pai Chi Li*
  • *Corresponding author for this work
  • National Taiwan University

Research output: Contribution to journalJournal Article peer-review

14 Scopus citations

Abstract

Conventional ultrasound (US) and photoacoustic (PA) multimodality imaging require the use of a US pulse for US data acquisition and a laser pulse for PA data acquisition. We propose a method for concurrent US and PA data acquisition with a single-laser pulse. A light-absorbing multilayer film that can generate a US pulse based on the thermoelastic effect is used. The selection of appropriate layer thickness, interlayer spacing, and absorption coefficient allows the spectral characteristics of the generated US signal to be adjusted so that it does not overlap with the spectrum of the PA signal generated by the light transmitting through the layer. Thus, the US signal and the PA signal can be generated, received, and separated by using a single-laser pulse combined with spectral filtering. This method is demonstrated using a multilayer film that generates US signals with a center frequency of 24.2 MHz and fractional bandwidth of 26.8%. The synthetic-aperture focusing technique is applied to improve the lateral resolution and the signal-to-noise ratio. A cyst-like phantom and a film phantom were used to demonstrate the feasibility of this method of concurrent PA-US imaging using single-laser pulses.

Original languageEnglish
Article number086004
JournalJournal of Biomedical Optics
Volume20
Issue number8
DOIs
StatePublished - 01 08 2015
Externally publishedYes

Bibliographical note

Publisher Copyright:
© 2015 Society of Photo-Optical Instrumentation Engineers (SPIE).

Keywords

  • laser-generated ultrasound
  • multimodality imaging
  • photoacoustic imaging

Fingerprint

Dive into the research topics of 'Concurrent photoacoustic-ultrasound imaging using single-laser pulses'. Together they form a unique fingerprint.

Cite this