Evaluation of the use of the fourth version FloTrac system in cardiac output measurement before and after cardiopulmonary bypass

Sheng Yi Lin, An Hsun Chou, Yung Fong Tsai, Su Wei Chang, Min Wen Yang, Pei Chi Ting, Chun Yu Chen*

*Corresponding author for this work

Research output: Contribution to journalJournal Article peer-review

13 Scopus citations

Abstract

The FloTrac system is a system for cardiac output (CO) measurement that is less invasive than the pulmonary artery catheter (PAC). The purposes of this study were to (1) compare the level of agreement and trending abilities of CO values measured using the fourth version of the FloTrac system (CCO-FloTrac) and PAC-originated continuous thermodilution (CCO-PAC) and (2) analyze the inadequate CO-discriminating ability of the FloTrac system before and after cardiopulmonary bypass (CPB). Fifty patients were included. After exclusion, 32 patients undergoing cardiac surgery with CPB were analyzed. All patients were monitored with a PAC and radial artery catheter connected to the FloTrac system. CO was assessed at 10 timing points during the surgery. In the Bland–Altman analysis, the percentage errors (bias, the limits of agreement) of the CCO-FloTrac were 61.82% (0.16, − 2.15 to 2.47 L min) and 51.80% (0.48, − 1.97 to 2.94 L min) before and after CPB, respectively, compared with CCO-PAC. The concordance rates in the four-quadrant plot were 64.10 and 62.16% and the angular concordance rates (angular mean bias, the radial limits of agreement) in the polar-plot analysis were 30.00% (17.62°, − 70.69° to 105.93°) and 38.63% (− 10.04°, − 96.73° to 76.30°) before and after CPB, respectively. The area under the receiver operating characteristic curve for CCO-FloTrac was 0.56, 0.52, 0.52, and 0.72 for all, ≥ ± 5, ≥ ± 10, and ≥ ± 15% CO changes (ΔCO) of CCO-PAC before CPB, respectively, and 0.59, 0.55, 0.49, and 0.46 for all, ≥ ± 5, ≥ ± 10, and ≥ ± 15% ΔCO of CCO-PAC after CPB, respectively. When CO < 4 L/min was considered inadequate, the Cohen κ coefficient was 0.355 and 0.373 before and after CPB, respectively. The accuracy, trending ability, and inadequate CO-discriminating ability of the fourth version of the FloTrac system in CO monitoring are not statistically acceptable in cardiac surgery.

Original languageEnglish
Pages (from-to)807-815
Number of pages9
JournalJournal of Clinical Monitoring and Computing
Volume32
Issue number5
DOIs
StatePublished - 01 10 2018

Bibliographical note

Publisher Copyright:
© 2017, Springer Science+Business Media B.V.

Keywords

  • Arterial waveform analysis
  • Cardiac output
  • Cardiopulmonary bypass
  • Cohen κ coefficient
  • FloTrac

Fingerprint

Dive into the research topics of 'Evaluation of the use of the fourth version FloTrac system in cardiac output measurement before and after cardiopulmonary bypass'. Together they form a unique fingerprint.

Cite this