Abstract
This paper reports a case involving failure of the polyethylene liner of a Mecron bipolar hip prosthesis with impingement of varus metal shell on the femoral stem neck resulting in deep defect and impending fracture of femoral stem components.
| Original language | English |
|---|---|
| Pages (from-to) | 131-134 |
| Number of pages | 4 |
| Journal | Journal of Musculoskeletal Research |
| Volume | 6 |
| Issue number | 2 |
| DOIs | |
| State | Published - 2002 |
| Externally published | Yes |
Keywords
- Bipolar hemiarthroplasty
- Component failure
- Femoral stem
- Revision joint arthroplasty