Abstract
High sensitivity detection on Faraday rotation and Verdet constant of ambient air under weak applied magnet field and single-pass configuration specimen is setup in which a two-frequency paired linear polarized interferometer (TPPI) coupled with balanced detector shown at shot-noise-limited detection is demonstrated. The Verdet constant of ambient air at 1.3× 10-6 rad/mT m and its sensitivity at 4.3× 10-8 rad/mT m were measured. Additionally, this method also is able to extend into a broad spectral range on Faraday rotation angle or Verdet constant measurement. Finally, the enhancement on detection sensitivity by integrating a Fabry-Ṕrot cavity into TPPI is discussed.