Abstract
This work introduces an electrical field-based biosensor for accurate quantification of histamine, featuring a polyvinyl chloride (PVC) membrane. The device employs a highly sensitive analog front-end circuit with an 8- \mu V low-drift voltage offset, enabling detection over a wide concentration range of 0.02–100 \mu M, achieving a sensitivity of 12.35 mV/ \mu M. When exposed to histamine in the presence of diamine oxidase (DAO), the sensor exhibited a pronounced signal change of 63%. Marked signal suppression occurs exclusively under severe interference conditions, specifically when the concentration ratio of histamine to putrescine, cadaverine, and tyramine reaches 1:20. Owing to its excellent sensitivity and selectivity, the proposed histamine biosensor is well-suited for rapid, on-site histamine detection in food and fish markets, offering strong potential for food safety monitoring applications.
| Original language | English |
|---|---|
| Pages (from-to) | 12704-12711 |
| Number of pages | 8 |
| Journal | IEEE Sensors Journal |
| Volume | 26 |
| Issue number | 8 |
| DOIs | |
| State | Published - 15 04 2026 |
| Externally published | Yes |
Bibliographical note
Publisher Copyright:© 2001-2012 IEEE.
Keywords
- Biosensor
- diamine oxidase (DAO)
- electrical field
- histamine
- readout
Fingerprint
Dive into the research topics of 'Investigation of Electrocatalysis for Electrical Field-Based Anti-Interference Histamine Biosensor With Readout Circuit'. Together they form a unique fingerprint.Cite this