Objective: This study evaluates the feasibility and efficacy of implanting a leadless pacemaker at the right atrial appendage (RAA) in a preclinical minipig model, aiming to address the limitations of atrial pacing with current leadless devices like the Medtronic Micra, which is typically used for right ventricular implantation. Methods: Four minipigs, each with a median body weight of 45.8 ± 10.0 kg, underwent placement of the Micra transcatheter pacing system (TPS) via the right femoral vein into the RAA apex. The pacing performance was assessed over 1-week (short-term) and 3-month (long-term) periods. Outcomes: The initial findings indicated successful implantation, with satisfactory intrinsic R-wave amplitudes and pacing threshold. In the following period, the sensitivity, threshold, and impedance were stable with time. Notably, upon explanation at 3 months, a deep myocardial penetration by the device was observed, necessitating a redesign for safe long-term use in a growing subject's heart. Conclusion: While initial results suggest that RAA apex placement of the Micra TPS is promising for potential inclusion in a dual-chamber pacing system, the issue of myocardial penetration highlights the need for device redesign to ensure safety and effectiveness in long-term applications.

Original languageEnglish
Pages (from-to)1124-1127
Number of pages4
JournalPACE - Pacing and Clinical Electrophysiology
Volume47
Issue number8
DOIs
StatePublished - 08 2024
Externally publishedYes

Bibliographical note

© 2024 The Author(s). Pacing and Clinical Electrophysiology published by Wiley Periodicals LLC.

Keywords

  • leadless pacemaker
  • right atrial appendage
  • transcatheter pacing system
  • Atrial Appendage/surgery
  • Animals
  • Swine
  • Swine, Miniature
  • Cardiac Pacing, Artificial/methods
  • Equipment Design
  • Pacemaker, Artificial
  • Feasibility Studies

Fingerprint

Dive into the research topics of 'Leadless pacemaker implementation at the right atrial appendage apex: An initial preclinical assessment'. Together they form a unique fingerprint.

Cite this