Optimal skin surface landmark for the SVC-RA junction in cancer patients requiring the implantation of permanent central venous catheters

J. H. Hsu, S. S. Wang, D. V. Lu, Kuang I. Cheng*, C. K. Wang, J. R. Wu

*Corresponding author for this work

Research output: Contribution to journalJournal Article peer-review

4 Scopus citations

Abstract

We compared four different skin surface landmarks, the lower margin of the right 2nd costo-sternal junction (point A); the upper margin of the right 3rd costo-sternal junction (point B); the lower margin of the right 3rd costo-sternal junction (point C); and a point 5 cm below the manubrio-sternal junction (point D), in 20 cancer patients undergoing insertion of permanent central venous catheters whose tips were placed near the superior vena cava - right atrium (SVC-RA) junction under transoesophageal echocardiography guidance. The landmark was satisfactory if it was located within 1 cm of the SVC-RA junction. Points C and D were closer to the SVC-RA junction than points A and B (p < 0.0001). However, point C had the highest incidence (C: 70%, A: 0%, B: 20%, D: 30%, p < 0.0001) of being within 1 cm of the SVC-RA junction.

Original languageEnglish
Pages (from-to)818-823
Number of pages6
JournalAnaesthesia
Volume62
Issue number8
DOIs
StatePublished - 08 2007
Externally publishedYes