Optimal skin surface landmark for the SVC-RA junction in cancer patients requiring the implantation of permanent central venous catheters

  • J. H. Hsu
  • , S. S. Wang
  • , D. V. Lu
  • , Kuang I. Cheng*
  • , C. K. Wang
  • , J. R. Wu
  • *Corresponding author for this work

Research output: Contribution to journalJournal Article peer-review

5 Scopus citations

Abstract

We compared four different skin surface landmarks, the lower margin of the right 2nd costo-sternal junction (point A); the upper margin of the right 3rd costo-sternal junction (point B); the lower margin of the right 3rd costo-sternal junction (point C); and a point 5 cm below the manubrio-sternal junction (point D), in 20 cancer patients undergoing insertion of permanent central venous catheters whose tips were placed near the superior vena cava - right atrium (SVC-RA) junction under transoesophageal echocardiography guidance. The landmark was satisfactory if it was located within 1 cm of the SVC-RA junction. Points C and D were closer to the SVC-RA junction than points A and B (p < 0.0001). However, point C had the highest incidence (C: 70%, A: 0%, B: 20%, D: 30%, p < 0.0001) of being within 1 cm of the SVC-RA junction.

Original languageEnglish
Pages (from-to)818-823
Number of pages6
JournalAnaesthesia
Volume62
Issue number8
DOIs
StatePublished - 08 2007
Externally publishedYes

UN SDGs

This output contributes to the following UN Sustainable Development Goals (SDGs)

  1. SDG 3 - Good Health and Well-being
    SDG 3 Good Health and Well-being

Fingerprint

Dive into the research topics of 'Optimal skin surface landmark for the SVC-RA junction in cancer patients requiring the implantation of permanent central venous catheters'. Together they form a unique fingerprint.

Cite this