Abstract
Objectives Evaluate the performance of Adaptive Iterative Dose Reduction 3D (AIDR 3D) and compare with filtered-back projection (FBP) regarding radiation dosage and image quality for an 80-kVp abdominal CT. Materials and methods An abdominal phantom underwent four CT acquisitions and reconstruction algorithms (FBP; AIDR 3D mild, standard and strong). Sixty-three patients underwent unenhanced liver CT with FBP and standard level AIDR 3D. Further post-acquisition reconstruction with strong level AIDR 3D was made. Patients were divided into two groups (< and ≧29 cm) based on the abdominal effective diameter (Deff) at T12 level. Quantitative (attenuation, noise, and signal-to-noise ratio) and qualitative (image quality, noise, sharpness, and artifact) analysis by two readers were assessed and the interobserver agreement was calculated. Results Strong level AIDR 3D reduced radiation dose by 72% in the phantom and 47.1% in the patient study compared with FBP. There was no difference in mean attenuations. Image noise was the lowest and signal-to-noise ratio the highest using strong level AIDR 3D in both patient groups. For Deff < 29 cm, image sharpness of FBP was significantly different from those of AIDR 3D (P < 0.05). For Deff ≧29 cm, image quality of AIDR 3D was significantly more favorable than FBP (P < 0.05). Interobserver agreement was substantial. Conclusions Integrated AIDR 3D allows for an automatic reduction in radiation dose and maintenance of image quality compared with FBP. Using AIDR 3D reconstruction, patients with larger abdomen circumference could be imaged at 80 kVp.
Original language | English |
---|---|
Pages (from-to) | 1666-1672 |
Number of pages | 7 |
Journal | European Journal of Radiology |
Volume | 85 |
Issue number | 9 |
DOIs | |
State | Published - 01 09 2016 |
Bibliographical note
Publisher Copyright:© 2016 Elsevier Ireland Ltd
Keywords
- Imaging
- Multidetector computed tomography
- Observer variation
- Phantoms
- Radiation dosage
- Radiographic image enhancement