Abstract
This in vitro experiment aimed to understand the difference in preload acting on an abutment screw under different angles of angulated screw-retained crown and the performance after cyclic loading. In total, thirty implants with angulated screw channel (ASC) abutments were divided into two parts. The first part consisted of three groups: a 0° access channel with a zirconia crown (ASC-0) (n = 5), a 15° access channel with a specially designed zirconia crown (sASC-15) (n = 5), and a 25° access channel with a specially designed zirconia crown (sASC-25) (n = 5). The reverse torque value (RTV) was measured at 0° for each specimen. The second part consisted of three groups: a 0° access channel with a zirconia crown (ASC-0) (n = 5); a 15° access channel with a zirconia crown (ASC-15) (n = 5), and a 25° access channel with a zirconia crown (ASC-25) (n = 5). The manufacturer’s recommended torque was applied to each specimen, and baseline RTV was measured before cyclic loading. Each ASC implant assembly was cyclically loaded at 0 to 40 N with 1 million cycles at 10 Hz. RTV was measured after cyclic loading. Kruskal–Wallis test and Jonckheere–Terpstra test were used for statistical analysis. All specimens were examined under a digital microscope and scanning electron microscope (SEM) to observe the wear of the screw head before and after the whole experiment. A significant difference in the different percentages of straight RTV (sRTV) between the three groups was found (p = 0.027). The angle of ASC to the different percentages of sRTV showed a significant linear trend (p = 0.003). No significant differences were found in RTV difference after cyclic loading among the ASC-0, ASC-15, and ASC-25 groups (p = 0.212). The ASC-25 group had the most serious degree of wear based on a digital microscope and SEM examination. The ASC angle will affect the actual preload acting on a screw: the larger the ASC angle, the smaller the preload. The performance of the angled ASC groups in RTV difference was comparable to that of 0° ASC after cyclic loading.
| Original language | American English |
|---|---|
| Article number | 124 |
| Journal | Journal of Functional Biomaterials |
| Volume | 14 |
| Issue number | 3 |
| DOIs | |
| State | Published - 24 02 2023 |
Bibliographical note
Publisher Copyright:© 2023 by the authors.
Keywords
- angulated screw channel
- angulated screw-retained crown
- cyclic loading
- implant
- mechanical stresses
- reverse torque value
Fingerprint
Dive into the research topics of 'Reverse Torque Value of Angulated Screw Channel Abutment before and after Cyclic Loading: An In Vitro Study.'. Together they form a unique fingerprint.Cite this
- APA
- Author
- BIBTEX
- Harvard
- Standard
- RIS
- Vancouver