Abstract
Percutaneous transluminal angioplasty (PTA) for thrombosed dialysis graft is both difficult and time-consuming that may increase radiation exposure. A predictor of PTA success and procedure time has not been identified yet. Sonography data and the PTA results of 88 grafts were reviewed retrospectively. The echogenicity of the majority of the intra-graft thrombus and the presence/absence of thrombus free space in the grafts were examined. The association between the sonography findings and PTA procedure success and procedure time was analyzed. The echogenicity of the intra-graft thrombus and presence/absence of thrombus free space were the major determinants of procedure success and procedure time. Higher intra-graft echogenicity and lack of thrombus free space were associated with more procedure failures (p < 0.01 vs. p = 0.04) and longer procedure times (p = 0.03 vs. p < 0.01). Thrombi from occluded dialysis grafts may differ in resistance to mechanical/pharmacological thrombolysis. Sonography characteristics of the intra-graft thrombus may help to differentiate them.
Original language | English |
---|---|
Pages (from-to) | 545-550 |
Number of pages | 6 |
Journal | Ultrasound in Medicine and Biology |
Volume | 38 |
Issue number | 4 |
DOIs | |
State | Published - 04 2012 |
Externally published | Yes |
Keywords
- Percutaneous transluminal angioplasty
- Prosthetic dialysis graft
- Sonography
- Thrombus