Survey of computed tomography scanners in Taiwan: Dose descriptors, dose guidance levels, and effective doses

H. Y. Tsai, C. J. Tung*, C. C. Yu, Y. S. Tyan

*Corresponding author for this work

Research output: Contribution to journalJournal Article peer-review

31 Scopus citations

Abstract

The IAEA and the ICRP recommended dose guidance levels for the most frequent computed tomography (CT) examinations to promote strategies for the optimization of radiation dose to CT patients. A national survey, including on-site measurements and questionnaires, was conducted in Taiwan in order to establish dose guidance levels and evaluate effective doses for CT. The beam quality and output and the phantom doses were measured for nine representative CT scanners. Questionnaire forms were completed by respondents from facilities of 146 CT scanners out of 285 total scanners. Information on patient, procedure, scanner, and technique for the head and body examinations was provided. The weighted computed tomography dose index (CTDIw), the dose length product (DLP), organ doses and effective dose were calculated using measured data, questionnaire information and Monte Carlo simulation results. A cost-effective analysis was applied to derive the dose guidance levels on CTDIw and DLP for several CT examinations. The mean effective dose±standard deviation distributes from 1.6±0.9 mSv for the routine head examination to 13±11 mSv for the examination of liver, spleen, and pancreas. The surveyed results and the dose guidance levels were provided to the national authorities to develop quality control standards and protocols for CT examinations.

Original languageEnglish
Pages (from-to)1234-1243
Number of pages10
JournalMedical Physics
Volume34
Issue number4
DOIs
StatePublished - 2007
Externally publishedYes

Fingerprint

Dive into the research topics of 'Survey of computed tomography scanners in Taiwan: Dose descriptors, dose guidance levels, and effective doses'. Together they form a unique fingerprint.

Cite this