Abstract
The IAEA and the ICRP recommended dose guidance levels for the most frequent computed tomography (CT) examinations to promote strategies for the optimization of radiation dose to CT patients. A national survey, including on-site measurements and questionnaires, was conducted in Taiwan in order to establish dose guidance levels and evaluate effective doses for CT. The beam quality and output and the phantom doses were measured for nine representative CT scanners. Questionnaire forms were completed by respondents from facilities of 146 CT scanners out of 285 total scanners. Information on patient, procedure, scanner, and technique for the head and body examinations was provided. The weighted computed tomography dose index (CTDIw), the dose length product (DLP), organ doses and effective dose were calculated using measured data, questionnaire information and Monte Carlo simulation results. A cost-effective analysis was applied to derive the dose guidance levels on CTDIw and DLP for several CT examinations. The mean effective dose±standard deviation distributes from 1.6±0.9 mSv for the routine head examination to 13±11 mSv for the examination of liver, spleen, and pancreas. The surveyed results and the dose guidance levels were provided to the national authorities to develop quality control standards and protocols for CT examinations.
| Original language | English |
|---|---|
| Pages (from-to) | 1234-1243 |
| Number of pages | 10 |
| Journal | Medical Physics |
| Volume | 34 |
| Issue number | 4 |
| DOIs | |
| State | Published - 2007 |
| Externally published | Yes |