Abstract
Background: Previous studies have shown that routine heated humidifier (HH) do not provide any benefit during continuous positive airway pressure (CPAP) titration if there are no significant naso-pharyngeal symptoms. In clinical practice, nasal diseases and upper airway symptoms are very common. This study investigates the effects of HH during CPAP titration in subjects with or without naso-pharyngeal symptoms. Methods: Fifty-two patients who received polysomnography with CPAP titration were randomly assigned to HH and non-HH groups. Their nasal cavity, pharynx, and naso-pharynx were evaluated before CPAP titration, and a questionnaire on subjective sensation, including naso-pharyngeal symptoms, willingness to further use CPAP, and sleep improvement, was used. Objective (e.g., leak, apnea-hypopnea index (AHI) reduction, and optimal CPAP pressure level) and subjective data were analyzed between the two groups. Results: In subjective sensation, the HH group did not have any benefit in further willingness to use CPAP and in sleep improvement, but had improved naso-pharyngeal symptoms (p = 0.043). There were no significant differences in leak, AHI reduction, and optimal CPAP pressure, even in patients with significant naso-pharyngeal symptoms. Conclusion: Routine use of HH is not necessary during CPAP titration regardless of naso-pharyngeal symptoms.
| Original language | English |
|---|---|
| Pages (from-to) | 133-138 |
| Number of pages | 6 |
| Journal | Sleep and Breathing |
| Volume | 17 |
| Issue number | 1 |
| DOIs | |
| State | Published - 03 2013 |
| Externally published | Yes |
Keywords
- Continuous positive airway pressure (CPAP) titration
- Humidifier
- Obstructive sleep apnea-hypopnea syndrome