摘要
This paper presents the performance of dynamic light scattering (DLS) and differential mobility analyzer (DMA) instruments in measuring nanoparticle size data using a gauge repeatability and reproducibility (GR&R) study based on the mixed model. The adequacy of the measurement process was evaluated using a measure of %GR&R with its confidence interval. The results show that the performance of the DMA instrument is superior to that of the DLS with regard to all variance components of a GR&R study. However, both instruments performed poorly in terms of %GR&R. Therefore, the part and measurement system require further improvement. Using these empirical study applications, this research promotes and maintains measurement quality through procedure analysis.
原文 | 英語 |
---|---|
期刊 | Journal of Testing and Evaluation |
卷 | 40 |
發行號 | 4 |
DOIs | |
出版狀態 | 已出版 - 07 2012 |
對外發佈 | 是 |