A novel biodegradable polycaprolactone fixator for osteosynthesis surgery of rib fracture: In vitro and in vivo study

  • Yi Hsun Yu
  • , Chin Lung Fan
  • , Yung Heng Hsu
  • , Ying Chao Chou
  • , Steve W.N. Ueng
  • , Shih Jung Liu*
  • *此作品的通信作者
  • Chang Gung Memorial Hospital
  • Chang Gung University

研究成果: 期刊稿件文章同行評審

13 引文 斯高帕斯(Scopus)

摘要

Osteosynthesis surgery for rib fractures is controversial and challenging. This study developed a noval poly(e(open)-caprolactone) (PCL)-based biodegradable "cable-tie" fixator for osteosynthesis surgery for rib fractures. A biodegradable fixator specifically for fractured ribs was designed and fabricated by a micro-injection molding machine in our laboratory. The fixator has three belts that could be passed through matching holes individually. The locking mechanism allows the belt movement to move in only one direction. To examine the in vitro biomechanical performance, ribs 3-7 from four fresh New Zealand rabbits were employed. The load to failure and stress-strain curve was compared in the three-point bending test among native ribs, titanium plate-fixed ribs, and PCL fixator-fixed ribs. In the in vivo animal study, the sixth ribs of New Zealand rabbits were osteotomized and osteosynthesis surgery was performed using the PCL fixator. Outcomes were assessed by monthly X-ray examinations, a final micro-computed tomography (CT) scan, and histological analysis. The experimental results suggested that the ribs fixed with the PCL fixator were significantly less stiffthan those fixed with titanium plates (p < 0.05). All ribs fixed with the PCL fixators exhibited union. The bridging callus was confirmed by gross, radiographic micro-three-dimensional (3D) CT, and histological examinations. In addition, there was no significant inflammatory response of the osteotomized ribs or the PCL-rib interface during application. The novel PCL fixator developed in this work achieves satisfactory results in osteosynthesis surgery for rib fractures, and may provide potential applications in other orthopedic surgeries.

原文英語
頁(從 - 到)7714-7722
頁數9
期刊Materials
8
發行號11
DOIs
出版狀態已出版 - 2015

文獻附註

Publisher Copyright:
© 2015 by the authors.

指紋

深入研究「A novel biodegradable polycaprolactone fixator for osteosynthesis surgery of rib fracture: In vitro and in vivo study」主題。共同形成了獨特的指紋。

引用此