Adaptive Iterative Dose Reduction 3D Integrated with Automatic Tube Current Modulation for CT Coronary Artery Calcium Quantification: Comparison to Traditional Filtered Back Projection in an Anthropomorphic Phantom and Patients

Ya Chun Tang, Yuan Chang Liu, Ming Yi Hsu, Hui Yu Tsai, Chien Ming Chen*

*此作品的通信作者

研究成果: 期刊稿件文章同行評審

8 引文 斯高帕斯(Scopus)

摘要

Rationale and Objectives: We aimed to evaluate integrated adaptive iterative dose reduction 3D (AIDR 3D) algorithm in automatic tube current modulation (ATCM) for the quantification of coronary artery calcium score (CACS) and cardiac risk stratification. Materials and Methods: A thoracic phantom with calcium inserts of known densities was scanned with filtered back projection (FBP) and AIDR 3D algorithms in small- and medium-sized phantoms. Twenty-four patients underwent two consecutive scans of CACS with FBP and AIDR 3D algorithms. The absolute Agatston score, Agatston score risk, volume score, and Agatston score percentile-based risk were compared, and concordance coefficients and agreement plots were made. Results: Agatston and volume scores were significantly different between the phantom sizes (P <.01). There were no significant differences in the Agatston scores between FBP and AIDR 3D for the medium phantoms (P =.25). In the patients, there were no significant differences in Agatston and volume scores between FBP and AIDR 3D (P =.06 and P =.09, respectively). The correlation coefficients of Agatston and volume scores with AIDR 3D were excellent compared to those of FBP. There were no significant differences in Agatston score risk and Agatston score percentile-based risk between FBP and AIDR 3D (P =.74 and P = 1, respectively). There was mean dose reduction of 57.8% ± 18.6% for AIDR 3D. Conclusion: The absolute Agatston score differed between FBP and AIDR 3D reconstructions. However, the cardiac risk categorizations of the two methods were comparable. An integrated AIDR 3D algorithm with automatic tube current modulation enables radiation dose savings at a consistent noise level without sacrificing CACS.

原文英語
頁(從 - 到)1010-1017
頁數8
期刊Academic Radiology
25
發行號8
DOIs
出版狀態已出版 - 08 2018

文獻附註

Publisher Copyright:
© 2018 The Association of University Radiologists

指紋

深入研究「Adaptive Iterative Dose Reduction 3D Integrated with Automatic Tube Current Modulation for CT Coronary Artery Calcium Quantification: Comparison to Traditional Filtered Back Projection in an Anthropomorphic Phantom and Patients」主題。共同形成了獨特的指紋。

引用此