TY - JOUR
T1 - Aerosol Delivery Using Jet Nebulizer and Vibrating Mesh Nebulizer during High Frequency Oscillatory Ventilation
T2 - An in Vitro Comparison
AU - Fang, Tien Pei
AU - Lin, Hui Ling
AU - Chiu, Shu Hua
AU - Wang, Szu Hui
AU - Diblasi, Robert M.
AU - Tsai, Ying Huang
AU - Fink, James B.
N1 - Publisher Copyright:
© Copyright 2016, Mary Ann Liebert, Inc.
PY - 2016/10
Y1 - 2016/10
N2 - Background: High frequency oscillatory ventilation (HFOV) is used in critically ill patients with severe hypoxemic respiratory failure. The purpose of this in vitro study was to determine the efficiency of aerosol delivery with different lung parameters during simulated neonatal, pediatric, and adult HFOV. Methods: Sensormedics 3100A/B ventilators were used to deliver infant, pediatric, and adult HFOV. Two types of aerosol generators were chosen for testing: 1) a continuous jet nebulizer (JN) with a unit-dose of 5.0 mg/2.5 mL salbutamol sulfate diluted into 4 mL, and 2) a vibrating mesh nebulizer (VMN) with salbutamol sulfate were run to completion of aerosol generation. Both aerosol devices were placed 1) between the ventilator circuit and the endotracheal tube (ETT) (proximal position); and 2) at the inlet of the heated humidifier (distal position) (n = 5). Drug was collected on a bacterial filter placed distal to the ETT, and the drug eluted and analyzed with a UV Spectrophotometer at 276 nm. T-test and ANOVA tests were used for comparison (p < 0.05). Results: The inhaled drug delivered by JN was 0%-0.6% of the nominal dose when placed at distal position, and 0%-3% at proximal position (p < 0.01), while the VMN was 0%-0.5% at distal and 8.6%-22.7% at proximal position (p < 0.01). Aerosol delivery during HFOV was greater with adult settings than pediatric and infant settings with VMN and JN (22.7%, 8.6%, and 17.4% respectively, p < 0.01). When the aerosol delivery device was placed at the distal position, negligible drug mass was observed (<0.5%), regardless of the nebulizer device used. Conclusions: During HFOV, aerosol delivery with the nebulizer placed at proximal was greater than placement distal from the ETT, with VMN delivering more drug than JN. The inhaled drug was delivery correlated positively with ETT size, MAP, and bias flow, and inversely proportional to power settings.
AB - Background: High frequency oscillatory ventilation (HFOV) is used in critically ill patients with severe hypoxemic respiratory failure. The purpose of this in vitro study was to determine the efficiency of aerosol delivery with different lung parameters during simulated neonatal, pediatric, and adult HFOV. Methods: Sensormedics 3100A/B ventilators were used to deliver infant, pediatric, and adult HFOV. Two types of aerosol generators were chosen for testing: 1) a continuous jet nebulizer (JN) with a unit-dose of 5.0 mg/2.5 mL salbutamol sulfate diluted into 4 mL, and 2) a vibrating mesh nebulizer (VMN) with salbutamol sulfate were run to completion of aerosol generation. Both aerosol devices were placed 1) between the ventilator circuit and the endotracheal tube (ETT) (proximal position); and 2) at the inlet of the heated humidifier (distal position) (n = 5). Drug was collected on a bacterial filter placed distal to the ETT, and the drug eluted and analyzed with a UV Spectrophotometer at 276 nm. T-test and ANOVA tests were used for comparison (p < 0.05). Results: The inhaled drug delivered by JN was 0%-0.6% of the nominal dose when placed at distal position, and 0%-3% at proximal position (p < 0.01), while the VMN was 0%-0.5% at distal and 8.6%-22.7% at proximal position (p < 0.01). Aerosol delivery during HFOV was greater with adult settings than pediatric and infant settings with VMN and JN (22.7%, 8.6%, and 17.4% respectively, p < 0.01). When the aerosol delivery device was placed at the distal position, negligible drug mass was observed (<0.5%), regardless of the nebulizer device used. Conclusions: During HFOV, aerosol delivery with the nebulizer placed at proximal was greater than placement distal from the ETT, with VMN delivering more drug than JN. The inhaled drug was delivery correlated positively with ETT size, MAP, and bias flow, and inversely proportional to power settings.
KW - Aerosol delivery
KW - breathing patterns
KW - high frequency oscillatory ventilation
KW - jet nebulizer
KW - vibrating mesh nebulizer
UR - http://www.scopus.com/inward/record.url?scp=84989930228&partnerID=8YFLogxK
U2 - 10.1089/jamp.2015.1265
DO - 10.1089/jamp.2015.1265
M3 - 文章
C2 - 26974175
AN - SCOPUS:84989930228
SN - 1941-2711
VL - 29
SP - 447
EP - 453
JO - Journal of Aerosol Medicine and Pulmonary Drug Delivery
JF - Journal of Aerosol Medicine and Pulmonary Drug Delivery
IS - 5
ER -