Beam profile monitoring system for proton therapy

Chili Ho*, Augustine E. Chen, Shujhen Dai, Minglee Chu, Chih Hsun Lin, Ping Kun Teng, Chung Hsiang Wang, Tsi Chian Chao, Chung Chi Lee, Chuang Jong Tung, Ting Shien Duh

*此作品的通信作者

研究成果: 圖書/報告稿件的類型會議稿件同行評審

摘要

An innovative design of 16-channel front-end electronics for ionization chamber in monitoring proton beam profile has been developed to feature simplicity, low cost, durability, and reliability without compromising its performances. All the parts are available off the shelf. The front-end electronics board is powered by USB connection, through which the data are transferred to the host computer for storage and further online analysis. The system has been studied with Geant4 MC [1] and tested with both electron and proton beams.

原文英語
主出版物標題IEEE Nuclear Science Symposuim and Medical Imaging Conference, NSS/MIC 2010
頁面440-442
頁數3
DOIs
出版狀態已出版 - 2010
事件2010 IEEE Nuclear Science Symposium, Medical Imaging Conference, NSS/MIC 2010 and 17th International Workshop on Room-Temperature Semiconductor X-ray and Gamma-ray Detectors, RTSD 2010 - Knoxville, TN, 美國
持續時間: 30 10 201006 11 2010

出版系列

名字IEEE Nuclear Science Symposium Conference Record
ISSN(列印)1095-7863

Conference

Conference2010 IEEE Nuclear Science Symposium, Medical Imaging Conference, NSS/MIC 2010 and 17th International Workshop on Room-Temperature Semiconductor X-ray and Gamma-ray Detectors, RTSD 2010
國家/地區美國
城市Knoxville, TN
期間30/10/1006/11/10

指紋

深入研究「Beam profile monitoring system for proton therapy」主題。共同形成了獨特的指紋。

引用此