Biomechanical comparison of different combinations of hook and screw in one spine motion unit - An experiment in porcine model

  • Chang Gung University

研究成果: 期刊稿件文章同行評審

15 引文 斯高帕斯(Scopus)

摘要

Background: The biomechanical performance of the hooks and screws in spinal posterior instrumentation is not well-characterized. Screw-bone interface failure at the uppermost and lowermost vertebrae is not uncommon. Some have advocated for the use of supplement hooks to prevent screw loosening. However, studies describing methods for combined hook and screw systems that fully address the benefits of these systems are lacking. Thus, the choice of which implant to use in a given case is often based solely on a surgeon's experience instead of on the biomechanical features and advantages of each device. Methods. We conducted a biomechanical comparison of devices instrumented with different combinations of hooks and screws. Thirty-six fresh low thoracic porcine spines were assigned to three groups (12 per group) according to the configuration used for of fixation: (1) pedicle screw; (2) lamina hook and (3) combination of pedicle screw and lamina hook. Axial pullout tests backward on transverse plane in the direction normal to the rods were performed using a material testing machine and a specially designed grip with self-aligned function. Results: The pullout force for the pedicle screws group was significantly greater than for the hooks and the combination (p < 0.05). However, no significant difference was found between the hooks and the combination (p > 0.05). Conclusions: Pedicle screws achieve the maximal pullout strength for spinal posterior instrumentation.

原文英語
文章編號197
期刊BMC Musculoskeletal Disorders
15
發行號1
DOIs
出版狀態已出版 - 09 06 2014

指紋

深入研究「Biomechanical comparison of different combinations of hook and screw in one spine motion unit - An experiment in porcine model」主題。共同形成了獨特的指紋。

引用此