Biomechanical evaluation of a modified dorsal double-plating fixation with adjustable joint and microthread designs for comminuted extra-articular distal radius fractures

  • Yu Hao Lin
  • , Alvin Chao Yu Chen
  • , Hsien Nan Kuo
  • , Tsung Chih Yu
  • , Ming Tsung Sun
  • , Chun Li Lin*
  • *此作品的通信作者
  • Chang Gung University
  • Chang Gung Memorial Hospital
  • Metal Industries Research and Development Center
  • National Yang Ming Chiao Tung University

研究成果: 期刊稿件文章同行評審

5 引文 斯高帕斯(Scopus)

摘要

This study evaluates the biomechanical strength of a modified dorsal double-plating (MDDP) internal fixation on artificial radius bones using the dynamic fatigue test. The plates are 1.6 mm thick and the upper L plate has an adjustable joint to allow the angular disposition of the bone plate to be changed to secure various bone fracture types. A dual-thread locking screw is used to enhance mechanical retention at the screw/bone interface. A finite element analysis is performed to examine the mechanical bone and fixation system response using a dual-microthread locking screw. An MDDP fixation is made with stainless steel by a manufacturer with ISO13485 quality management systems. Eighteen radius Sawbones were randomly placed into three groups and cut to a standard length of 2.5 cm from the articular surface to form 0 °, 30°, and 60° fracture configurations according to the A2 classification in the AO surgery reference. The adjustable joint in the L plates is adjusted and fixed using a fastener connection to have a 0°, 30°, or 60° configuration. Plates and screws are then positioned on the dorso-ulnar side for L plates and the dorso-radial side for I plates to create an angle of 90°. The specimens were subjected to oscillating loads of 10 to 150 N at 5 Hz for 20000 cycles. The average stiffness values after 20000 cycles were 425.7, 461.1, and 532.1 N/mm for the 0 °, 30°, and 60° constructs, respectively. No difference in stiffness was found for constructs with a given angle throughout the 20000 cycles of testing (p > 0.05). However, significant differences (p < 0.05) in stiffness were found between constructs with different angles at each 500-cycle interval. The lack of gross construct failure during cyclic testing and the reasonable stiffness prove that MDDP internal fixation is sufficiently stable to support restricted postoperative loads.

原文英語
頁(從 - 到)29-34
頁數6
期刊Journal of Medical and Biological Engineering
33
發行號1
DOIs
出版狀態已出版 - 2013

指紋

深入研究「Biomechanical evaluation of a modified dorsal double-plating fixation with adjustable joint and microthread designs for comminuted extra-articular distal radius fractures」主題。共同形成了獨特的指紋。

引用此