Biomechanical evaluation of an orthodontic mini-implant with plastic removal cap: An in-vitro experimental testing

研究成果: 圖書/報告稿件的類型 › 會議稿件 › 同儕審查

1 引文 斯高帕斯(Scopus)

摘要

This study evaluates the biomechanical interactions of a mini-implant using a plastic revolving cap (PRC) with translation/rotation features for optional orthodontic traction. An orthodontic mini-implant and the PRC consisting of a hexagon connection onto mini-implant with 60 degree switching unit and an extended arm to provide orthodontic wire tied at different positions. The PRC removal force was measured by pull-out testing. The PRC removal force remained larger than three times the finger pulling force (9.3N) after 5 repeated removal tests. The results for the PRC resistant testing showed that the PRC rotational resistant force (20.31±0.83N) is larger than the maximum traction force (about 4.9N) for orthodontic treatment. The mini-implant used with PRC can provide translation and rotation features to change the angles and directions of orthodontic tractions for most effective anchorage preparation under safety consideration.

原文英語
主出版品物標題2013 35th Annual International Conference of the IEEE Engineering in Medicine and Biology Society, EMBC 2013
頁面3183-3185
頁數3
DOIs
出版品狀態已出版 - 2013
事件2013 35th Annual International Conference of the IEEE Engineering in Medicine and Biology Society, EMBC 2013 - Osaka, 日本
持續時間: 03 07 2013 → 07 07 2013

出版品系列

名字Proceedings of the Annual International Conference of the IEEE Engineering in Medicine and Biology Society, EMBS
ISSN(列印)1557-170X

Conference

Conference2013 35th Annual International Conference of the IEEE Engineering in Medicine and Biology Society, EMBC 2013
國家/地區日本
城市Osaka
期間03/07/13 → 07/07/13

指紋

深入研究「Biomechanical evaluation of an orthodontic mini-implant with plastic removal cap: An in-vitro experimental testing」主題。共同形成了獨特的指紋。

引用此