Comparison of two detector systems for cone beam CT small animal imaging - A preliminary study

  • Yang Meng*
  • , Chris C. Shaw
  • , Xinming Liu
  • , Mustafa C. Altunbas
  • , Tianpeng Wang
  • , Lingyun Chen
  • , Shu Ju Tu
  • , S. Cheenu Kappadath
  • , Chao Jen Lai
  • *此作品的通信作者
  • University of Texas Health Science Center at Houston

研究成果: 圖書/報告稿件的類型會議稿件同行評審

4 引文 斯高帕斯(Scopus)

摘要

Purpose: To compare two detector systems - one based on the charge-coupled device (CCD) and image amplifier, the other based on a-Si/CsI flat panel, for cone beam computed-tomography (CT) imaging of small animals. A high resolution, high framing rate detector system for the cone beam CT imaging of small animals was developed. The system consists of a 2048×3072×12 bit CCD optically coupled to an image amplifier and an x-ray phosphor screen. The CCD has an intrinsic pixel size of 12 μm but the effective pixel size can be adjusted through the magnification adjustment of the optical coupling systems. The system is used in conjunction with an x-ray source and a rotating stage for holding and rotating the scanned object in the cone beam CT imaging experiments. The advantages of the system include but are not limited to the ability to adjust the effective pixel size and to achieve extremely high spatial resolution and temporal resolution. However, the need to use optical coupling compromises the detective quanta efficiency (DQE) of the system. In this paper, the imaging characteristics of the system were presented and compared with those of an a-Si/CsI flat-panel detector system.

原文英語
主出版物標題Medical Imaging 2006
主出版物子標題Physics of Medical Imaging
DOIs
出版狀態已出版 - 2006
對外發佈
事件Medical Imaging 2006: Physics of Medical Imaging - San Diego, CA, 美國
持續時間: 12 02 200616 02 2006

出版系列

名字Progress in Biomedical Optics and Imaging - Proceedings of SPIE
6142 III
ISSN(列印)1605-7422

Conference

ConferenceMedical Imaging 2006: Physics of Medical Imaging
國家/地區美國
城市San Diego, CA
期間12/02/0616/02/06

指紋

深入研究「Comparison of two detector systems for cone beam CT small animal imaging - A preliminary study」主題。共同形成了獨特的指紋。

引用此