Complications of liner locking system in MicroStructured Omnifit acetabular components: A radiographic evaluation of 887 hips followed for 5-10 years

  • Po Cheng Lee*
  • , Chun-Hsiung Shih
  • , Wen Lin Yen
  • , Wen E. Yang
  • , Yuan Kun Tu
  • , Ching Lung Tai
  • *此作品的通信作者
  • Chang Gung Memorial Hospital

研究成果: 期刊稿件文章同行評審

4 引文 斯高帕斯(Scopus)

摘要

We evaluated 887 hips in 672 patients with uncemented MicroStructured Omnifit acetabular components for liner locking complications. We found 2 types of radiographic signs of liner locking system complications in 7 hips, developing between 2 and 4 years postoperatively. The incidence of liner locking system complications was 0.8% using this modular acetabular component. We recommend that a patient who has received a total hip arthroplasty including a MicroStructured Omnifit acetabular component should be monitored frequently for radiographic signs of liner locking system complications, especially with a polyethylene thickness of less than 8 mm.