跳至主導覽 跳至搜尋 跳過主要內容

Computer-assisted fluoroscopic navigation of pedicle screw insertion: An in vivo feasibility study

  • Tsai Sheng Fu
  • , Li Huei Chen
  • , Chak Bor Wong
  • , Po Liang Lai
  • , Tsung Ting Tsai
  • , Chi Chieh Niu
  • , Wen Jer Chen
  • Chang Gung Memorial Hospital

研究成果: 期刊稿件 › 文獻綜述 › 同行評審

60 引文 斯高帕斯(Scopus)

摘要

Background: Accurate placement of pedicle screws is difficult. Patients and methods: We evaluated the feasibility and accuracy of pedicle screw insertion assisted by a real-time, 2-dimensional (2D) image-guided navigation system in 12 patients who underwent thoraco-lumbar and/or lumbar stabilization. 66 pedicle screws were inserted either by senior spine surgeons or residents. The accuracy of positioning of the screws was evaluated using postoperative plain radiographs and thin-cut CT. Results: 61 of the 66 screws were inserted successfully. 5 screw insertions showed structural violations: 4 on the medial and 1 on the lateral pedicle wall. The accuracy was higher in the sagittal plane than in the axial plain. There was no difference between the surgical error rates caused by the senior surgeons and the residents. Interpretation: Using computer-assisted 2D fluoroscopic image navigation, it is possible to achieve reliable and accurate pedicle screw insertion during low thoracic and lumbar spinal surgery.

原文英語
頁(從 - 到)730-735
頁數6
期刊Acta Orthopaedica Scandinavica
卷75
發行號6
DOIs
出版狀態已出版 - 12 2004
對外發佈是