Considering angle selection when using ultrasound electrode displacement elastography to evaluate radiofrequency ablation of tissues

Jingjing Xia, Qiang Li, Pin Yu Chen, Zhuhuang Zhou, Chiao Yin Wang, Hao Li Liu, Jianfu Teng, Po Hsiang Tsui*

*此作品的通信作者

研究成果: 期刊稿件文章同行評審

1 引文 斯高帕斯(Scopus)

摘要

Percutaneous radiofrequency ablation (RFA) is a minimally invasive treatment to thermally destroy tumors. Ultrasound-based electrode-displacement elastography is an emerging technique for evaluating the region of RFA-induced lesions. The angle between the imaging probe and the RFA electrode can influence electrode-displacement elastography when visualizing the ablation zone. We explored the angle effect on electrode-displacement elastography to measure the ablation zone. Phantoms embedded with meatballs were fabricated and then ablated using an RFA system to simulate RFA-induced lesions. For each phantom, a commercial ultrasound scanner with a 7.5 MHz linear probe was used to acquire raw image data at different angles, ranging from 30° to 90° at increments of 10°, to construct electrode-displacement images and facilitate comparisons with tissue section images. The results revealed that the ablation regions detected using electrode-displacement elastography were highly correlated with those from tissue section images when the angle was between 30° and 60°. However, the boundaries of lesions were difficult to distinguish, when the angle was larger than 60°. The experimental findings suggest that angle selection should be considered to achieve reliable electrode-displacement elastography to describe ablation zones.

原文英語
文章編號764320
期刊BioMed Research International
2014
DOIs
出版狀態已出版 - 2014

指紋

深入研究「Considering angle selection when using ultrasound electrode displacement elastography to evaluate radiofrequency ablation of tissues」主題。共同形成了獨特的指紋。

引用此