Design and test of a monolithic ultrasound-image-guided HIFU device using annular CMUT rings

Mengli Wang*, Jingkuang Chen, Xiaoyang Cheng, Jui Ching Cheng, Pai Chi Li

*此作品的通信作者

研究成果: 期刊稿件會議文章同行評審

8 引文 斯高帕斯(Scopus)

摘要

This paper describes the design, fabrication, and characterization of a CMUT-based therapeutic ultrasound chip with built-in ultrasound imager for real-time monitoring of the object being operated on. Multiple concentric high-power (inner) CMUT rings and an annular imager CMUT array (outmost) comprising of 48 or 64 elements are integrated on a silicon substrate measuring 2mm × 2mm for simultaneous ultrasonic ablation/stimulation and imaging. The polysilicon membrane thickness and gap height of the high-power CMUT devices are 1.3μm and 0.35 μm, respectively, while these of the imager CMUT integrated on the same substrate are 1.0μm and 0.18μm, respectively. The thicker membrane and higher gap of the high-power CMUTs make them capable of delivering highpressure ultrasound, while the thin membrane and lower gap of imager CMUTs improves the receiver sensitivity in pulse-echo imaging. In order to maximize the overall membrane displacement of the high-power CMUT ring in transmission mode, the ring is designed as a one-chamber swimming-ring structure instead of being divided into multiple sub-chambers, like imager arrays. The high-power CMUT rings were successfully used for heating the liver tissue of a pig, creating a 2.5°C temperature increase after 6 minutes of ultrasound irradiation. Preliminary B-mode imaging using the CMUT imager element on this imager-guided therapeutic chip was also demonstrated.

原文英語
文章編號4803667
頁(從 - 到)459-462
頁數4
期刊Proceedings - IEEE Ultrasonics Symposium
DOIs
出版狀態已出版 - 2008
事件2008 IEEE International Ultrasonics Symposium, IUS 2008 - Beijing, 中國
持續時間: 02 11 200805 11 2008

指紋

深入研究「Design and test of a monolithic ultrasound-image-guided HIFU device using annular CMUT rings」主題。共同形成了獨特的指紋。

引用此