TY - JOUR
T1 - Development of high throughput optical sensor array for on-line pH monitoring in micro-scale cell culture environment
AU - Wu, Min Hsien
AU - Lin, Jr Lung
AU - Wang, Junbo
AU - Cui, Zhanfeng
AU - Cui, Zheng
PY - 2009
Y1 - 2009
N2 - On-line pH detection of cell culture environment is necessary in a bioprocess or tissue engineering. Devices by means of electrochemical mechanisms for this purpose have been reported to be less suitable compared with optical-based sensing principles. More recently, some non-invasive optical sensing systems have been proposed for online pH monitoring of cell culture environment. However, these devices are not for multi-target pH monitoring purpose, and are large in scale and thus not appropriate for the pH monitoring at a micro scale such as in microbioreactor or microfluidic-based cell culture platform. To tackle these issues, an optical fiber sensor array for on-line pH monitoring was proposed using microfluidic technology. The working principle is based on the optical absorption of phenol red normally contained in culture medium. Different from other device of the similar working principle, the proposed device requires less liquid volume (less than 0.8 μl), is non-invasive, and particularly can be configured as an array for high throughput pH monitoring. The present device has been optimized for the shape of detection chamber in a microfluidic chip with the aid of computational fluid dynamics (CFD) simulation, to avoid flow dead zone and thus to reduce the response time of detection. Both simulation and experimental results revealed that the design of oval detection chamber (axis, 1.5 and 2.0 mm) can considerably reduce the response time. Preliminary test has proved that the optical pH detection device is able to detect pH with average detection sensitivity of 0.83 V/pH in the pH range of 6.8-7.8, which is normally experienced in mammalian cell culture.
AB - On-line pH detection of cell culture environment is necessary in a bioprocess or tissue engineering. Devices by means of electrochemical mechanisms for this purpose have been reported to be less suitable compared with optical-based sensing principles. More recently, some non-invasive optical sensing systems have been proposed for online pH monitoring of cell culture environment. However, these devices are not for multi-target pH monitoring purpose, and are large in scale and thus not appropriate for the pH monitoring at a micro scale such as in microbioreactor or microfluidic-based cell culture platform. To tackle these issues, an optical fiber sensor array for on-line pH monitoring was proposed using microfluidic technology. The working principle is based on the optical absorption of phenol red normally contained in culture medium. Different from other device of the similar working principle, the proposed device requires less liquid volume (less than 0.8 μl), is non-invasive, and particularly can be configured as an array for high throughput pH monitoring. The present device has been optimized for the shape of detection chamber in a microfluidic chip with the aid of computational fluid dynamics (CFD) simulation, to avoid flow dead zone and thus to reduce the response time of detection. Both simulation and experimental results revealed that the design of oval detection chamber (axis, 1.5 and 2.0 mm) can considerably reduce the response time. Preliminary test has proved that the optical pH detection device is able to detect pH with average detection sensitivity of 0.83 V/pH in the pH range of 6.8-7.8, which is normally experienced in mammalian cell culture.
KW - Microfluidic devices
KW - On-line monitoring
KW - Optical fiber
KW - Optical sensor
KW - pH
UR - http://www.scopus.com/inward/record.url?scp=59949095449&partnerID=8YFLogxK
U2 - 10.1007/s10544-008-9233-0
DO - 10.1007/s10544-008-9233-0
M3 - 文章
C2 - 18830696
AN - SCOPUS:59949095449
SN - 1387-2176
VL - 11
SP - 265
EP - 273
JO - Biomedical Microdevices
JF - Biomedical Microdevices
IS - 1
ER -