摘要
Purpose: The aim of this study was to understand whether the pitch relationship between micro and macro thread designs with a parametrical relationship in a dual-thread mini-implant can improve primary stability.
Materials and Methods: Three types of mini-implants consisting of single-thread (ST) (0.75 mm pitch in whole length), dual-thread A (DTA) with double-start 0.375 mm pitch, and dual-thread B (DTB) with single-start 0.2 mm pitch in upper 2-mm micro thread region for performing insertion and pull-out testing. Histomorphometric analysis was performed in these specimens in evaluating peri-implant bone defects using a non-contact vision measuring system.
Results: The maximum inserted torque (Tmax) in type DTA was found to be the smallest significantly, but corresponding values found no significant difference between ST and DTB. The largest pull-out strength (Fmax) in the DTA mini-implant was found significantly greater than that for the ST mini-implant regardless of implant insertion orientation. Miniimplant engaged the cortical bone well as observed in ST and DTA types.
Conclusion: Dual-thread miniimplant with correct micro thread pitch (parametrical relationship with macro thread pitch) in the cortical bone region can improve primary stability and enhanced mechanical retention.
原文 | 英語 |
---|---|
頁(從 - 到) | 653-658 |
頁數 | 6 |
期刊 | Implant Dentistry |
卷 | 23 |
發行號 | 6 |
DOIs | |
出版狀態 | 已出版 - 11 12 2014 |