Do dual-thread orthodontic mini-implants improve bone/tissue mechanical retention?

Yang Sung Lin, Yau Zen Chang, Jian Hong Yu, Chun Li Lin*

*此作品的通信作者

研究成果: 期刊稿件文章同行評審

5 引文 斯高帕斯(Scopus)

摘要

Purpose: The aim of this study was to understand whether the pitch relationship between micro and macro thread designs with a parametrical relationship in a dual-thread mini-implant can improve primary stability.

Materials and Methods: Three types of mini-implants consisting of single-thread (ST) (0.75 mm pitch in whole length), dual-thread A (DTA) with double-start 0.375 mm pitch, and dual-thread B (DTB) with single-start 0.2 mm pitch in upper 2-mm micro thread region for performing insertion and pull-out testing. Histomorphometric analysis was performed in these specimens in evaluating peri-implant bone defects using a non-contact vision measuring system.

Results: The maximum inserted torque (Tmax) in type DTA was found to be the smallest significantly, but corresponding values found no significant difference between ST and DTB. The largest pull-out strength (Fmax) in the DTA mini-implant was found significantly greater than that for the ST mini-implant regardless of implant insertion orientation. Miniimplant engaged the cortical bone well as observed in ST and DTA types.

Conclusion: Dual-thread miniimplant with correct micro thread pitch (parametrical relationship with macro thread pitch) in the cortical bone region can improve primary stability and enhanced mechanical retention.

原文英語
頁(從 - 到)653-658
頁數6
期刊Implant Dentistry
23
發行號6
DOIs
出版狀態已出版 - 11 12 2014

文獻附註

Publisher Copyright:
Copyright © 2014 by Lippincott Williams & Wilkins.

指紋

深入研究「Do dual-thread orthodontic mini-implants improve bone/tissue mechanical retention?」主題。共同形成了獨特的指紋。

引用此