摘要
In 1998, an epithermal neutron test beam was designed and constructed at the Tsing Hua Open-Pool Reactor (THOR) for the purpose of preliminary dosimetric experiments in boron neutron capture therapy (BNCT). A new epithermal neutron beam was designed at this facility, and is currently under construction, with clinical trials targeted in late 2004. Depth dose-rate distributions for the THOR BNCT test beam have been measured by means of activation foil and dual ion chamber techniques. Neutron and structure-induced gamma spectra measured at the test beam exit were configured into a source function for the Monte Carlo-based treatment planning code NCTPlan. Dose-rate scaling factors (DRSFs) were determined to normalize computationally derived dose-rate distributions with experimental measurements in corresponding mathematical and physical phantoms, and to thus enable accurate treatment planning using the NCTPlan code. A similar approach will be implemented in characterizing the new THOR epithermal beam in preparation for clinical studies. This paper reports the in-phantom calculated and experimental dosimetry comparisons and derived DRSFs obtained with the THOR test beam.
原文 | 英語 |
---|---|
頁(從 - 到) | 881-885 |
頁數 | 5 |
期刊 | Applied Radiation and Isotopes |
卷 | 61 |
發行號 | 5 |
DOIs | |
出版狀態 | 已出版 - 11 2004 |
對外發佈 | 是 |