TY - JOUR
T1 - Effect of Nebulizer Location and Spontaneous Breathing on Aerosol Delivery during Airway Pressure Release Ventilation in Bench Testing
AU - Ge, Hui Qing
AU - Wang, Ji Mei
AU - Lin, Hui Ling
AU - Fink, James B.
AU - Luo, Ronghua
AU - Xu, Peifeng
AU - Ying, Kejing
N1 - Publisher Copyright:
© 2018, Mary Ann Liebert, Inc., publishers.
PY - 2019/2
Y1 - 2019/2
N2 - Background: Airway pressure release ventilation (APRV) maintains a sustained airway pressure over a large proportion of the respiratory cycle, and has a long inspiratory time at high pressure. The purpose of this study was to determine the influence of the APRV with and without spontaneous breathing on albuterol aerosol delivery with a continuous vibrating-mesh nebulizer (VMN) placed at different positions on an adult lung model of invasive mechanical ventilation. Methods: An adult lung model was assembled by connecting a ventilator with a dual-limb circuit to an 8-mm inner diameter endotracheal tube (ETT) and collecting filter attached to a test lung with set compliance of 0.1 L/cmH 2 O and resistance of 0.5 cmH 2 O/(L·s). Four ventilator modes were compared: pressure control ventilation (PCV) with no bias flow, PCV with bias flow of 6 L/min (PCV BF6 ), APRV with no spontaneous breaths (APRV), and APRV with spontaneous breath trigger (APRVs). Peak inspiratory pressure, peak end-expiratory pressure, aerosol dose, and nebulization time were similar for all modes. The VMN was placed (1) between Y-piece and inspiratory limb, (2) at the gas outlet of a heated humidifier, and (3) at the gas inlet of a heated humidifier. Albuterol sulfate (5 mg/2.5 mL) was administered with each run and collected on a filter distal to the ETT. Deposited drug was eluted from each filter (purified water) and analyzed by UV spectrophotometry at 276 nm. Analysis of variance [general linear model (GLM) multivariate] was performed using the linear model of multiple variables, significance at p < 0.05. Results: Albuterol (in micrograms, mean ± standard deviation) delivered was higher with VMN placed at the gas inlet of the humidifier with each mode of ventilation (p < 0.01). APRVs has the highest albuterol delivery followed by PCV, PCV BF6 , and APRV (1706.2 ± 60.9 μg vs. 1490.6 ± 61.1 μg vs. 1182.3 ± 61.4 μg vs. 1153.1 ± 99.7 μg, respectively, p < 0.001). The minute volume was positively correlated with the inhaled albuterol dose. Conclusions: Spontaneous breathing increased the albuterol delivery during APRV, compared with APRV alone and PCV modes. Placing the nebulizer proximal to the ventilator was more efficient for all modes tested.
AB - Background: Airway pressure release ventilation (APRV) maintains a sustained airway pressure over a large proportion of the respiratory cycle, and has a long inspiratory time at high pressure. The purpose of this study was to determine the influence of the APRV with and without spontaneous breathing on albuterol aerosol delivery with a continuous vibrating-mesh nebulizer (VMN) placed at different positions on an adult lung model of invasive mechanical ventilation. Methods: An adult lung model was assembled by connecting a ventilator with a dual-limb circuit to an 8-mm inner diameter endotracheal tube (ETT) and collecting filter attached to a test lung with set compliance of 0.1 L/cmH 2 O and resistance of 0.5 cmH 2 O/(L·s). Four ventilator modes were compared: pressure control ventilation (PCV) with no bias flow, PCV with bias flow of 6 L/min (PCV BF6 ), APRV with no spontaneous breaths (APRV), and APRV with spontaneous breath trigger (APRVs). Peak inspiratory pressure, peak end-expiratory pressure, aerosol dose, and nebulization time were similar for all modes. The VMN was placed (1) between Y-piece and inspiratory limb, (2) at the gas outlet of a heated humidifier, and (3) at the gas inlet of a heated humidifier. Albuterol sulfate (5 mg/2.5 mL) was administered with each run and collected on a filter distal to the ETT. Deposited drug was eluted from each filter (purified water) and analyzed by UV spectrophotometry at 276 nm. Analysis of variance [general linear model (GLM) multivariate] was performed using the linear model of multiple variables, significance at p < 0.05. Results: Albuterol (in micrograms, mean ± standard deviation) delivered was higher with VMN placed at the gas inlet of the humidifier with each mode of ventilation (p < 0.01). APRVs has the highest albuterol delivery followed by PCV, PCV BF6 , and APRV (1706.2 ± 60.9 μg vs. 1490.6 ± 61.1 μg vs. 1182.3 ± 61.4 μg vs. 1153.1 ± 99.7 μg, respectively, p < 0.001). The minute volume was positively correlated with the inhaled albuterol dose. Conclusions: Spontaneous breathing increased the albuterol delivery during APRV, compared with APRV alone and PCV modes. Placing the nebulizer proximal to the ventilator was more efficient for all modes tested.
KW - Aerosol delivery
KW - airway pressure release ventilation
KW - bench testing
KW - spontaneous breath
KW - ventilation mode
KW - vibrating-mesh nebulizer
UR - http://www.scopus.com/inward/record.url?scp=85060819647&partnerID=8YFLogxK
U2 - 10.1089/jamp.2018.1457
DO - 10.1089/jamp.2018.1457
M3 - 文章
C2 - 30199313
AN - SCOPUS:85060819647
SN - 1941-2711
VL - 32
SP - 34
EP - 39
JO - Journal of Aerosol Medicine and Pulmonary Drug Delivery
JF - Journal of Aerosol Medicine and Pulmonary Drug Delivery
IS - 1
ER -