Experimental validation of Monte Carlo simulation model for proton range verification using an in-beam dual-head PET system

  • Qin Nan Gao
  • , Fang Jing Li
  • , Meng Wei Ho
  • , Fang Hsin Chen
  • , Vanny Maranatha Sihotang
  • , Eng Yen Huang
  • , Kun Ju Lin
  • , Chi Shiun Chiang
  • , Hsin Hon Lin*
  • *此作品的通信作者
  • National Tsing Hua University
  • Chang Gung University
  • Chang Gung Memorial Hospital
  • New York Proton Center

研究成果: 期刊稿件 › 文章 › 同行評審

1 引文 斯高帕斯(Scopus)

摘要

Proton therapy offers highly localized dose distributions, but its clinical potential is limited by uncertainties in proton range. In-beam positron emission tomography (PET) imaging of proton-induced β+-emitting isotopes provides a promising solution for real-time range verification. This study presents a comprehensive Monte Carlo (MC) simulation model for dual-head PET (DHPET)-based range verification using the GATE platform, with experimental validation performed at Kaohsiung Chang Gung Memorial Hospital, Taiwan. Measurements were conducted using homogeneous phantoms irradiated by proton beams at various energies, and PET images were acquired with a Sumitomo DHPET system. The simulation workflow was divided into two stages: proton-induced β+ isotope production and PET image acquisition. Three nuclear models were implemented and compared: GEANT4's built-in theoretical model (QGSP_BIC), EXFOR-based cross sections, and the latest Nuclear Data Sheets (NDS) dataset. The MC model was validated against RayStation-calculated dose distributions and PET measurements from both point and flood sources. Simulated dose distributions showed excellent agreement with RayStation, with a mean range deviation of ±0.2 mm and detector response deviation within ±0.5%. For PET activity profiles, the NDS dataset achieved the closest match to experimental data, EXFOR showed moderate agreement, while QGSP_BIC, despite discrepancies in activity shape, provided superior distal activity range estimates in gel-water phantoms. Under optimal conditions, proton range verification accuracy using the proposed MC simulation model and updated nuclear cross-section models could reach sub-millimeter precision (within ±1 mm). This work establishes a validated MC framework for in-beam DHPET-based range verification and emphasizes the pivotal role of nuclear cross-section selection in optimizing PET-based range estimation.

原文英語
文章編號113263
期刊Radiation Physics and Chemistry
卷238
DOIs
出版狀態已出版 - 01 2026

文獻附註

Publisher Copyright:
© 2025 Elsevier Ltd.

指紋

深入研究「Experimental validation of Monte Carlo simulation model for proton range verification using an in-beam dual-head PET system」主題。共同形成了獨特的指紋。

引用此