Failure analysis of broken pedicle screws on spinal instrumentation

  • Chen Sheng Chen
  • , Wen Jer Chen
  • , Cheng Kung Cheng*
  • , Shyh Hua Eric Jao
  • , Shan Chang Chueh
  • , Chang Chih Wang
  • *此作品的通信作者
  • National Yang Ming Chiao Tung University
  • Chang Gung Memorial Hospital
  • Industrial Technology Research Institute of Taiwan

研究成果: 期刊稿件 › 文章 › 同行評審

135 引文 斯高帕斯(Scopus)

摘要

Revised spinal surgery is needed when there is a broken pedicle screw in the patient. This study investigated the pedicle screw breakage by conducting retrieval analyses of broken pedicle screws from 16 patients clinically and by performing stress analyses in the posterolateral fusion computationally using finite element (FE) models. Fracture surface of screws was studied by scanning electron microscope (SEM). The FE model of the posterolateral fusion with the screw showed that screws on the caudal side had larger axial stress than those on the cephalic side, supporting the clinical findings that 75% of the patients had the screw breakage on the caudal side. SEM fractography showed that all broken screws exhibited beach marks or striations on the fractured surface, indicating fatigue failure. Screws of patients with spinal fracture showed fatigue striations and final ductile fracture around the edge. Among the 16 patients who had broken pedicle screws 69% of them achieved bone union in the bone graft, showing that bone union in the bone graft did not warrant the prevention of screw breakage.

原文英語
頁(從 - 到)487-496
頁數10
期刊Medical Engineering and Physics
卷27
發行號6
DOIs
出版狀態已出版 - 07 2005
對外發佈是

指紋

深入研究「Failure analysis of broken pedicle screws on spinal instrumentation」主題。共同形成了獨特的指紋。

引用此