In vitro evaluation of a vibrating-mesh nebulizer repeatedly use over 28 days

Hui Ling Lin*, Chi Shuo Chen, James B. Fink, Guo Hao Lee, Chun Wei Huang, Jui Chi Chen, Zi Yi Chiang

*此作品的通信作者

研究成果: 期刊稿件文章同行評審

15 引文 斯高帕斯(Scopus)

摘要

This in vitro study evaluates the performance of a disposable vibrating-mesh nebulizer when used for 28 days. A lung model was used to simulate the breathing pattern of an adult with chronic obstructive pulmonary disease. The vibrating-mesh nebulizer was used for three treatments/day over 28 days without cleaning after each test. Results showed that the inhaled drug dose was similar during four weeks of use (p = 0.157), with 16.73 ± 4.46% at baseline and 15.29 ± 2.45%, 16.21 ± 2.21%, 17.56 ± 1.98%, and 17.13 ± 1.81%, after the first, second, third, and fourth weeks, respectively. The particle size distribution, residual drug volume, and nebulization time remained similar across four weeks of use (p = 0.110, p = 0.763, and p = 0.573, respectively). Mesh was inspected using optical microscopy and showed that approximately 50% of mesh pores were obscured after 84 runs, and light penetration through the aperture plate was significantly reduced after the 21st use (p < 0.001) with no correlation to nebulizer performance. We conclude that the vibrating-mesh nebulizer delivered doses of salbutamol solution effectively over four weeks without cleaning after each use even though the patency and clarity of the aperture plate were reduced by the first week of use.

原文英語
文章編號0971
頁(從 - 到)1-9
頁數9
期刊Pharmaceutics
12
發行號10
DOIs
出版狀態已出版 - 10 2020

文獻附註

Publisher Copyright:
© 2020 by the authors. Licensee MDPI, Basel, Switzerland.

指紋

深入研究「In vitro evaluation of a vibrating-mesh nebulizer repeatedly use over 28 days」主題。共同形成了獨特的指紋。

引用此