Interference screw versus Endoscrew fixation for anterior cruciate ligament reconstruction: A biomechanical comparative study in sawbones and porcine knees

Chu Chih Hung, Wen Chuan Chen, Chan Tsung Yang, Cheng Kung Cheng, Chih Hwa Chen, Yu Shu Lai*

*此作品的通信作者

研究成果: 期刊稿件文章同行評審

18 引文 斯高帕斯(Scopus)

摘要

Interference screw fixation is one of the most common methods for ligament reconstruction. Although the advantages and clinical outcomes of this procedure have been widely reported, post-surgical complications often arise. The purpose of this study was to evaluate a new femoral fixation device, the Endoscrew, for anterior cruciate ligament (ACL) reconstruction. We performed a mechanical test in accordance with American Society for Testing and Materials (ASTM) standards and an invitro biomechanical study. An axial pullout test was conducted to evaluate the mechanical properties of the new device and the interference screw when implanted in solid rigid polyurethane foam test blocks. The biomechanical test used porcine femora to evaluate the initial fixation strength between these two implants. The maximum pullout force of the interference screw group [722.05±130.49N (N)] was significantly greater (p<0.01) than the Endoscrew group (440.79±26.54N) when implanted in polyurethane foam 320 kg/m3 density. With polyurethane foam 160 kg/m3 density, the maximum pullout forces were (242.61±37.36N) (p<0.001) and (99.33±30.01N) for the interference screw group and Endoscrew group, respectively. In the invitro mechanical study, the Endoscrew (646.39±72.38N) required a significantly greater ultimate load prior to failure (p<0.05) when compared with the interference screw (489.72±138.64N). With regard to pullout stiffness, there was no statistically significant difference (p<0.13) between the Endoscrew group (99.15±12.16N/mm) and the interference screw group (87.96±11.12N/mm). The cyclic stiffness was also not significantly different (p<0.44) between the Endoscrew group (93.09±16.07N/mm) and the interference screw group (85.78±14.76N/mm). The axial pullout test showed that the strength of the Endoscrew was close to the fixation strength required for daily activities, but it is inappropriate for use in osteoporotic patients. The invitro study showed that the Endoscrew has a superior initial mechanical fixation over the biodegradable interference screw.

原文英語
頁(從 - 到)82-90
頁數9
期刊Journal of Orthopaedic Translation
2
發行號2
DOIs
出版狀態已出版 - 04 2014
對外發佈

指紋

深入研究「Interference screw versus Endoscrew fixation for anterior cruciate ligament reconstruction: A biomechanical comparative study in sawbones and porcine knees」主題。共同形成了獨特的指紋。

引用此