摘要
In this study, we proposed a silicon nanobelt field-effect transistor (FET) to detect prostate specific antigen (PSA). The nanobelt FET device displayed n-channel depletion characteristics. The immobilization of prostate specific antibody (anti-PSA) molecules was attached onto the nanobelt FET surface by using the aldehyde groups of glutaraldehyde linked to the amino groups of 3-aminopropyltriethoxysilane (APTES). The shift in the drain current vs time curves of a nanobelt FET biosensor revealed that the electrical signal had a logarithmic relationship with respect to the concentration of the PSA, and detection capability was estimated in the 5 pg/mL level. To enhance the sensitivity of a nanobelt FET biosensor, this biosensor was designed by inserting arginine molecules between glutaraldehyde and APTES. Therefore, the detection capability of the developed sensor was extended to 50 fg/mL. Also, the relationship between the current shift and the logarithm of PSA concentration was exhibited linearity in the range 50 fg/mL - 500 pg/mL. The excellent electrical results of this label-free PSA nanobelt FET biosensor suggested that such biosensor might be potentially useful tools for biological research and future prostate cancer screening.
| 原文 | 英語 |
|---|---|
| 頁(從 - 到) | 4432-4442 |
| 頁數 | 11 |
| 期刊 | International Journal of Electrochemical Science |
| 卷 | 7 |
| 發行號 | 5 |
| DOIs | |
| 出版狀態 | 已出版 - 2012 |
UN SDG
此研究成果有助於以下永續發展目標
-
SDG3 健康與福祉
指紋
深入研究「Label-free detection of prostate specific antigen using a silicon nanobelt field-effect transistor」主題。共同形成了獨特的指紋。引用此
- APA
- Author
- BIBTEX
- Harvard
- Standard
- RIS
- Vancouver