In this research work, we present a flexible iridium oxide (IrOx) extended-gate field-effect transistor (EGFET) biosensor for label-free detection of the epidermal growth factor receptor (EGFR) biomarker. IrOx was employed as the sensing membrane material due to its excellent electrochemical properties, high conductivity, stability, biocompatibility, and ability to facilitate redox reactions. The IrOx film was deposited on a flexible polyimide (PI) substrate through a sol-gel process and characterized using atomic force microscopy and X-ray photoelectron spectroscopy. As a pH sensor, the IrOx EGFET exhibited a super-Nernstian sensitivity of 68.54 mV/pH with high linearity, a low hysteresis of ∼ 6 mV, and a drift rate of 0.50 mV/h at pH 7 over 12 h. For EGFR detection, the IrOx surface was functionalized with 3-aminopropyltriethoxysilane and glutaraldehyde to immobilize anti-EGFR antibodies. The EGFR sensor demonstrated a wide linear detection range from 1 to 1000 ng/mL with a sensitivity of 11.78 mV/log(ng/mL) and a linearity of 99.8%, making it suitable for clinical EGFR monitoring in cancer patients. The flexible IrOx EGFET shows promise for reliable label-free detection of EGFR.