Mass accuracy improvement of reversed-phase liquid chromatography/electrospray ionization mass spectrometry based urinary metabolomic analysis by post-run calibration using sodium formate cluster ions

  • Chiun Gung Juo*
  • , Chien Lun Chen
  • , Shiang Ting Lin
  • , Shu Hsuan Fu
  • , Yi Ting Chen
  • , Yu Sun Chang
  • , Jau Song Yu
  • *此作品的通信作者
  • Chang Gung University
  • Chang Gung Memorial Hospital

研究成果: 期刊稿件 › 文章 › 同行評審

16 引文 斯高帕斯(Scopus)

摘要

RATIONALE Typically, a batch metabolomics analysis using liquid chromatography/electrospray ionization time-of-flight mass spectrometry (LC/ESI-TOF MS) takes 2 to 3 days. However, the mass accuracy - which has an important influence on metabolite identification - can drift by as much as about 17 ppm in such a time period. In an untargeted urinary metabolomics analysis by reversed-phase liquid chromatography (RPLC)/ESI-MS, the signals of sodium formate cluster ions were detected at the column-washing step. The cluster ions were used to calibrate the mass spectrometer for more accurate detection. METHODS The spectra were calibrated post-run by the sodium formate cluster ions, which were used as the internal standard, in order to improve the mass accuracy. RESULTS In the analysis of urine samples, we calibrated the spectra acquired by the micrOTOF with the sodium cluster ions. In positive mode ESI, the average errors of these cluster ions were improved to ±0.48 ppm and in negative mode ESI, to ±0.94 ppm after calibration. The mass accuracy remained within ±0.01 ppm over the duration of 6.25 days. An error window of 4 ppm appears to be suitable for metabolite identification when using post-calibration. CONCLUSIONS The results showed that sodium formate cluster ions could be utilized for the calibration of LC/ESI-TOF MS and the average instrumental errors could be maintained at low levels for long-term analyses. This method could be applied not only to urine sample, but also to low sodium samples, such as saliva, by dissolving the sample in 1μM sodium formate solution. This method provides a good solution for accurate mass detection of metabolomic analysis.

原文英語
頁(從 - 到)1813-1820
頁數8
期刊Rapid Communications in Mass Spectrometry
卷28
發行號16
DOIs
出版狀態已出版 - 30 08 2014

文獻附註

Publisher Copyright:
Copyright © 2014 John Wiley & Sons, Ltd. Copyright © 2014 John Wiley & Sons, Ltd.

指紋

深入研究「Mass accuracy improvement of reversed-phase liquid chromatography/electrospray ionization mass spectrometry based urinary metabolomic analysis by post-run calibration using sodium formate cluster ions」主題。共同形成了獨特的指紋。

引用此