Measurement of polyethylene wear in cementless total hip arthroplasty

  • Chun Hsiung Shih*
  • , Po Cheng Lee
  • , Jian Horng Chen
  • , Ching Lung Tai
  • , Li Fan Chen
  • , J. Shih Shyn Wu
  • , Walter H. Chang
  • *此作品的通信作者
  • Chang Gung Memorial Hospital
  • Chang Gung University
  • Chung Yuan Christian University
  • National Chung Hsing University

研究成果: 期刊稿件文章同行評審

57 引文 斯高帕斯(Scopus)

摘要

We made a clinical study of polyethylene wear in 240 hips of 187 patients having primary total hip arthroplasties from 1989 to 1990, using uncemented Osteonics components, with a head size of 26 mm. We excluded cups with anteversion of over 20°and measured linear wear by a new method using a digitiser and special software of our design. Follow-up was from two to five years (mean 4.3). The mean age at operation was 50.3 years, with more men than women (1.4:1). The mean linear wear per year was 0.15mm; this did not increase with the longevity of the prosthesis (p = 0.54). In 59 hips showing evidence of osteolysis, the mean linear wear rate was significantly higher at 0.23 mm/year (p < 0.001). The mean linear wear rate also correlated significantly with age at the time of operation (p = 0.008), but we found no significant correlations with body-weight, gender, aetiology of the disease, thickness of polyethylene, or cup position. Our new method of measurement is time-saving and reproducible. The results confirm the greater rate of linear wear of polyethylene in patients showing osteolysis and in those who are younger.

原文英語
頁(從 - 到)361-365
頁數5
期刊Journal of Bone and Joint Surgery - Series B
79
發行號3
DOIs
出版狀態已出版 - 05 1997

指紋

深入研究「Measurement of polyethylene wear in cementless total hip arthroplasty」主題。共同形成了獨特的指紋。

引用此