T1 - Smart artificial cervical intervertebral disc system with a 3 × 3 PANI:PEO piezoresistive pressure sensor array and self-designed readout circuit for real-time intradiscal pressure monitoring
AU - Wang, Jer Chyi
AU - Li, Ying Ching
AU - Chen, Kuan Lin
AU - Lin, Hung Pin
AU - Chen, Min Wei
AU - Kao, Shao Ku
AU - Chang, Yau Zen
AU - Lu, Yu Jen
N1 - Publisher Copyright:
© 2026 Published by Elsevier B.V.
PY - 2026/12/16
Y1 - 2026/12/16
N2 - Cervical disc replacement is increasingly implemented to restore segmental motion after disc degeneration. However, the postoperative intradiscal biomechanics of the implanted prosthesis remain unobservable between follow-up visits, leaving early failure modes such as subsidence, migration, endplate microfracture, and facet overloading undetected until they become radiographically or symptomatically evident. Herein, a smart artificial cervical intervertebral disc is proposed, which integrates a flexible 3 × 3 polyaniline:polyethylene oxide (PANI:PEO) piezoresistive pressure sensor array with a self-designed 0.18-µm CMOS readout application-specific integrated circuit (ASIC) for real-time intradiscal pressure monitoring. The sensor array exhibited a strictly monotonic piezoresistive response up to 1000 kPa with a pixel sensitivity of 0.034–0.040 kPa−1, thus fully encompassing the physiological cervical loading range. The ASIC directly resolved gigaohm-level impedances (0.05–5 GΩ) with one order of magnitude above Wheatstone-bridge-based architectures and delivered a 7-bit on-chip digital output through a resistance-to-voltage converter, voltage-controlled oscillator, frequency-to-digital converter, and amplitude-shift-keying modulator within a total die area of 0.703 × 0.863 mm2 at 79 MHz for 117.81 mW. The fully integrated prosthesis was implanted in a human-shaped head-and-neck phantom and successfully resolved the characteristic load-transfer patterns across the three principal anatomical planes, capturing anterior/posterior peaks of 90–94 kPa under sagittal flexion/extension, ipsilateral peaks of up to 101 kPa under coronal lateral bending, and concentric central-core stress of 110–118 kPa under transverse axial rotation. The proposed platform successfully converts a passive prosthesis into a continuously self-reporting sentinel and lays the groundwork for intelligent and self-monitoring cervical disc implants.
AB - Cervical disc replacement is increasingly implemented to restore segmental motion after disc degeneration. However, the postoperative intradiscal biomechanics of the implanted prosthesis remain unobservable between follow-up visits, leaving early failure modes such as subsidence, migration, endplate microfracture, and facet overloading undetected until they become radiographically or symptomatically evident. Herein, a smart artificial cervical intervertebral disc is proposed, which integrates a flexible 3 × 3 polyaniline:polyethylene oxide (PANI:PEO) piezoresistive pressure sensor array with a self-designed 0.18-µm CMOS readout application-specific integrated circuit (ASIC) for real-time intradiscal pressure monitoring. The sensor array exhibited a strictly monotonic piezoresistive response up to 1000 kPa with a pixel sensitivity of 0.034–0.040 kPa−1, thus fully encompassing the physiological cervical loading range. The ASIC directly resolved gigaohm-level impedances (0.05–5 GΩ) with one order of magnitude above Wheatstone-bridge-based architectures and delivered a 7-bit on-chip digital output through a resistance-to-voltage converter, voltage-controlled oscillator, frequency-to-digital converter, and amplitude-shift-keying modulator within a total die area of 0.703 × 0.863 mm2 at 79 MHz for 117.81 mW. The fully integrated prosthesis was implanted in a human-shaped head-and-neck phantom and successfully resolved the characteristic load-transfer patterns across the three principal anatomical planes, capturing anterior/posterior peaks of 90–94 kPa under sagittal flexion/extension, ipsilateral peaks of up to 101 kPa under coronal lateral bending, and concentric central-core stress of 110–118 kPa under transverse axial rotation. The proposed platform successfully converts a passive prosthesis into a continuously self-reporting sentinel and lays the groundwork for intelligent and self-monitoring cervical disc implants.
KW - Application-specific integrated circuit (ASIC)
KW - Artificial intervertebral disc
KW - Frequency-to-digital converter
KW - Implantable sensor
KW - Piezoresistive pressure sensor array
KW - Polyaniline:polyethylene oxide (PANI:PEO)
UR - https://www.scopus.com/pages/publications/105053836453
U2 - 10.1016/j.sna.2026.118641
DO - 10.1016/j.sna.2026.118641
M3 - 文章
AN - SCOPUS:105053836453
SN - 0924-4247
VL - 412
JO - Sensors and Actuators A: Physical
JF - Sensors and Actuators A: Physical
M1 - 118641
ER -