Sphenoidal EEG recording using acupuncture needle electrode in complex partial seizure

N. S. Chu*, C. L. Wu, T. S. Tseng, L. L. Kuo

*此作品的通信作者

研究成果: 期刊稿件文章同行評審

10 引文 斯高帕斯(Scopus)

摘要

Sphenoidal EEG recording using an uninsulated acupuncture needle electrode were performed in 41 patients with or suspected of complex partial seizures of temporal lobe origin. The anterior temporal spikes were detected by the routine EEG in 17 patients (41%) and by the acupuncture sphenoidal needle in 29 patients (70%). The anterior temporal spikes recorded by the acupuncture needle were almost identical in configuration, amplitude and distribution to those recorded by conventional wire or insulated needle sphenoidal electrodes. The sequence in the frequency of spike detection by these 3 types of sphenoidal electrode were SP1-2, T1-2, F7-8 and A1-2 locations. The spikes of maximal amplitude were most frequently recorded by the SP electrode followed by the T1-2 electrode. The placement of the disposable acupuncture needle was simple and safe. Patients experienced minimal discomfort or pain that lasted at most 0.5 h. No complications occurred. The records were generally free of artifacts. It is concluded that the acupuncture needle can be used as sphenoidal electrode in outpatient EEG recording for the diagnosis of complex partial seizures of anterior temporal origin.

原文英語
頁(從 - 到)119-126
頁數8
期刊Electroencephalography and Clinical Neurophysiology