Eight fresh porcine lumbar spines received a posterior instrumentation at L4-L5 using pedicle screw-rod system. Each specimen was tested utilizing laminectomies of varying extent. Group A (Integrity) preserved the spinous process and interspinous ligament; Group B (Partial laminectomy) removed the inferior portion of L4 spinous process and preserved the interspinous ligament of L3-L4; Group C (Complete laminectomy) removed the entire L4 spinous process. Hydraulic testing machine was used to generate an increasing moment up to 8400 N mm in flexion and extension. The intervertebral displacement on the superior adjacent disc between L3-L4 was measured using an extensometer. Under extension, no significant difference in the intervertebral displacement was observed among three different models of laminectomy. However, under flexion, the intervertebral displacement on adjacent disc with complete laminectomy was statistically larger than those of integrity and partial laminectomies (P = 0.000976 and P = 0.0363, respectively). No difference was found between integrity and partial laminectomy groups (P > 0.05). This study implies that an instrumented spine with integrity of posterior complex is less likely to develop adjacent instability than a spine with destruction of the anchoring point for supraspinous ligament.

原文英語
頁(從 - 到)261-267
頁數7
期刊Bio-Medical Materials and Engineering
16
發行號4
出版狀態已出版 - 2006

指紋

深入研究「The effect of interspinous ligament integrity on adjacent segment instability after lumbar instrumentation and laminectomy - An experimental study in porcine model」主題。共同形成了獨特的指紋。

引用此