TY - JOUR
T1 - The Effects of Inspiratory Flows, Inspiratory Pause, and Suction Catheter on Aerosol Drug Delivery with Vibrating Mesh Nebulizers During Mechanical Ventilation
AU - Lin, HL
AU - Fink, James B
AU - Li, Jie
N1 - Publisher Copyright:
Copyright 2024, Mary Ann Liebert, Inc., publishers
PY - 2024/6/1
Y1 - 2024/6/1
N2 -
Background:
Some experts recommend specific ventilator settings during nebulization for mechanically ventilated patients, such as inspiratory pause, high inspiratory to expiratory ratio, and so on. However, it is unclear whether those settings improve aerosol delivery. Thus, we aimed to evaluate the impact of ventilator settings on aerosol delivery during mechanical ventilation (MV).
Methods:
Salbutamol (5.0 mg/2.5 mL) was nebulized by a vibrating mesh nebulizer (VMN) in an adult MV model. VMN was placed at the inlet of humidifier and 15 cm away from the Y-piece of the inspiratory limb. Eight scenarios with different ventilator settings were compared with endotracheal tube (ETT) connecting 15 cm from the Y-piece, including tidal volumes of 6-8 mL/kg, respiratory rates of 12-20 breaths/min, inspiratory time of 1.0-2.5 seconds, inspiratory pause of 0-0.3 seconds, and bias flow of 3.5 L/min. In-line suction catheter was utilized in two scenarios. Delivered drug distal to the ETT was collected by a filter, and drug was assayed by an ultraviolet spectrophotometry (276 nm).
Results:
Compared to the use of inspiratory pause, the inhaled dose without inspiratory pause was either higher or similar across all ventilation settings. Inhaled dose was negatively correlated with inspiratory flow with VMN placed at 15 cm away from the Y-piece (
r
s
= -0.68,
p < 0.001) and at the inlet of humidifier (
r
s
= -0.83,
p < 0.001). The utilization of in-line suction catheter reduced inhaled dose, regardless of the ventilator settings and nebulizer placements.
Conclusions:
When VMN was placed at the inlet of humidifier, directly connecting the Y-piece to ETT without a suction catheter improved aerosol delivery. In this configuration, the inhaled dose increased as the inspiratory flow decreased, inspiratory pause had either no or a negative impact on aerosol delivery. The inhaled dose was greater with VMN placed at the inlet of humidifier than 15 cm away the Y-piece.
AB -
Background:
Some experts recommend specific ventilator settings during nebulization for mechanically ventilated patients, such as inspiratory pause, high inspiratory to expiratory ratio, and so on. However, it is unclear whether those settings improve aerosol delivery. Thus, we aimed to evaluate the impact of ventilator settings on aerosol delivery during mechanical ventilation (MV).
Methods:
Salbutamol (5.0 mg/2.5 mL) was nebulized by a vibrating mesh nebulizer (VMN) in an adult MV model. VMN was placed at the inlet of humidifier and 15 cm away from the Y-piece of the inspiratory limb. Eight scenarios with different ventilator settings were compared with endotracheal tube (ETT) connecting 15 cm from the Y-piece, including tidal volumes of 6-8 mL/kg, respiratory rates of 12-20 breaths/min, inspiratory time of 1.0-2.5 seconds, inspiratory pause of 0-0.3 seconds, and bias flow of 3.5 L/min. In-line suction catheter was utilized in two scenarios. Delivered drug distal to the ETT was collected by a filter, and drug was assayed by an ultraviolet spectrophotometry (276 nm).
Results:
Compared to the use of inspiratory pause, the inhaled dose without inspiratory pause was either higher or similar across all ventilation settings. Inhaled dose was negatively correlated with inspiratory flow with VMN placed at 15 cm away from the Y-piece (
r
s
= -0.68,
p < 0.001) and at the inlet of humidifier (
r
s
= -0.83,
p < 0.001). The utilization of in-line suction catheter reduced inhaled dose, regardless of the ventilator settings and nebulizer placements.
Conclusions:
When VMN was placed at the inlet of humidifier, directly connecting the Y-piece to ETT without a suction catheter improved aerosol delivery. In this configuration, the inhaled dose increased as the inspiratory flow decreased, inspiratory pause had either no or a negative impact on aerosol delivery. The inhaled dose was greater with VMN placed at the inlet of humidifier than 15 cm away the Y-piece.
KW - in-line suction catheter
KW - inspiratory flow
KW - inspiratory pause
KW - mechanical ventilation
KW - nebulization
KW - vibrating mesh nebulizer
KW - Suction
KW - Vibration
KW - Humans
KW - Administration, Inhalation
KW - Tidal Volume
KW - Equipment Design
KW - Catheters
KW - Intubation, Intratracheal/instrumentation
KW - Inhalation
KW - Time Factors
KW - Respiration, Artificial/instrumentation
KW - Aerosols
KW - Adult
KW - Albuterol/administration & dosage
KW - Bronchodilator Agents/administration & dosage
KW - Nebulizers and Vaporizers
KW - Drug Delivery Systems/instrumentation
UR - http://www.scopus.com/inward/record.url?scp=85189878190&partnerID=8YFLogxK
U2 - 10.1089/jamp.2023.0026
DO - 10.1089/jamp.2023.0026
M3 - Journal Article
C2 - 38563958
SN - 1941-2711
VL - 37
SP - 125
EP - 131
JO - Journal of Aerosol Medicine and Pulmonary Drug Delivery
JF - Journal of Aerosol Medicine and Pulmonary Drug Delivery
IS - 3
ER -