Three-dimensional dose distribution in contrast-enhanced digital mammography using Gafchromic XR-QA2 films: Feasibility study

  • Yi Shuan Hwang
  • , Yu Ying Lin
  • , Yun Chung Cheung
  • , Hui Yu Tsai*
  • *此作品的通信作者
  • Chang Gung Memorial Hospital
  • Chang Gung University

研究成果: 期刊稿件文章同行評審

11 引文 斯高帕斯(Scopus)

摘要

This study was aimed to establish three-dimensional dose distributions for contrast-enhanced digital mammography (CEDM) using self-developed Gafchromic XR-QA2 films. Dose calibration and distribution evaluations were performed on a full-field digital mammography unit with dual energy (DE) contrast-enhanced option. Strategy for dose calibration of films in the DE mode was based on the data obtained from common target/filter/kVp combinations used clinically and the dose response model modified from Rampado's model. Dose derived from films were also verified by measured data from an ionization chamber. The average difference of dose was 8.9% in the dose range for clinical uses. Three-dimensional dose distributions were estimated using triangular acrylic phantom equipped with the mammography system. Five pieces of film sheets were separately placed between the acrylic slabs to evaluate the dose distribution at different depths. After normalizing the dose in each pixel to the maximum dose at the top-center position of the acrylic, normalized dose distribution for transverse, coronal and sagittal planes, could thus be obtained. The depth dose distribution evaluated in this study may further serve as a reference for evaluating the patient glandular dose at different depths based on the entrance exposure information.

原文英語
頁(從 - 到)204-207
頁數4
期刊Radiation Physics and Chemistry
104
DOIs
出版狀態已出版 - 11 2014

指紋

深入研究「Three-dimensional dose distribution in contrast-enhanced digital mammography using Gafchromic XR-QA2 films: Feasibility study」主題。共同形成了獨特的指紋。

引用此